User Equation
Input example
To input the equation for ΔE*ab
as a User Equation, it would be:
SQRT(POW([DL]+POW[DA]+POW[DB])
Input Format
A User Equation is defined using numbers and strings for measurement value variables and functions.
• Parameter strings are explained in Input Format below.
• The total length of a User Equation definition including numbers and string parameters is 200 characters.
Numerical constants
Numbers can be input as constants using 0 to 9 and the decimal point.
Although there is no limit on the number of decimal places that can be input, the number of significant digits for
calculations is 5 places.
Measurement value variables
Common variables
• Calculations will be performed using the SCI/SCE setting of the current screen. In the SCI screen, SCI values will
be used and in the SCE screen, SCE values will be used.
Variable
Measurement value
[L]
L*
[DL]
ΔL*
[C]
C*
[DC]
ΔC*
[X]
X
[DX]
ΔX
[SX]
x
[DSX]
Δx
[DE]
ΔE*ab
[DE00]
ΔE00
[DEWI]
ΔWI E313-73
Variable
Measurement value
[A]
a*
[DA]
Δa*
[H]
H
[DH]
ΔH*
[Y]
Y
[DY]
ΔY
[SY]
y
[DSY]
Δy
[CMC]
CMC
[MI]
M1
UE-2
Variable
Measurement value
[B]
b*
[DB]
Δb*
[Z]
Z
[DZ]
ΔZ
[DE94]
ΔE*94
[WIE]
WI E313-73