Measuring range:
Resolution:
Accuracy:
Ambient temperature:
Humidity:
Protection class:
Dimensions/weight:
Display:
Power supply:
Interface:
10
Measuring range - differential pressure: -0.05 kPa to 200 kPa
Max. static excess pressure:
Measuring range – temperature:
Differential pressure:
Flow:
Temperature:
Differential pressure:
Flow:
Temperature
Working temperature:
Storage:
Attention: In case of risk of frost, no water may remain in the
measuring head
In this case, the measuring head and the measuring hoses have to
be drained completely.
Max. relative humidity of 90% non-condensing
Body IP52
Keyboard IP54
Flow meter:
Measuring head:
LCD display with backlighting
4 rechargeable NiMh batteries
or by using enclosed power pack 230 V AC 50/60 Hz
or by using 4 commercial Mignon batteries 1.5 V
Attention: In case of battery operation, the power pack must not be
connected
USB
2000 kPa
-20°C to +120°C
0.01 kPa
0.0001 l/s
0.1°C
up to 10 kPa ± 0.1 kPa
10 to 200 kPa 1% of measured value
0.01 l/s
± 1°C
0°C to 40°C
-20°C to +60°C
160 x 63 x 40 mm, weight 470 g
130 x 70 x 70 mm, weight 1240 g