Introduction

Safe-Air Testers are designed to provide a quick and easy method for carrying out accurate periodic
testing of breathing-air supplies. From the touch screen display the appropriate breathing-air standard
can be selected for your test.
The F6001, is designed specifically to test high pressure air breathing air cylinders and high pressure
charging systems up to 300bar.
It is supplied in a hard weatherproof case with storage provision for a wide range of accessories and
equipment.
The test is carried out using built in electrochemical cells for carbon monoxide and the oxygen and an
infrared LED sensor for the carbon dioxide. The moisture is measured with an electronic dewpoint
meter and the instrument also records test pressure and ambient temperature. For measuring oil
aerosols Draeger Impactors are connected through a test port and this can also be used for testing a
number of other Draeger chemical reagent tubes detailed below. The Draeger Impactors are supplied
in packs of 10 and show the degree of oil contamination on a screen.
Calibration and Warranty
Safe-Air Testers leave our factory with a 12-month warranty and calibration certificate. Our standard
turnaround on annual calibration is 10-15 working days providing there is no major damage that
requires an extensive rebuild. Note:- Please download all stored data before returning the Tester to
Factair.
Temperature Parameters
Storage: -10/+50°C
This tester is calibrated for the following approved reagent tubes/Impactor.
Test
Oil (Impactor)
Oil
(chemical
tube)*
Nitrous Fumes
Sulphur Dioxide
Polytest
 The tester can also be used with the old type Draeger reagent tube for oil, ref.6728371. (Refer to
Appendix 2 at the rear of the manual for instructions.)
IMPORTANT – IT IS RECOMMENDED THAT YOUR SAFE-AIR TESTER IS RETURNED FOR
RECALIBRATION AND SERVICING WITHIN 12 MONTHS FROM THE ISSUE DATE OF ITS CALIBRATION
Operating Range: +5/+35°C
Draeger ref.
8103560
reagent
6728371
NO
8103661
x
SO
6728491
2
CH28401
Factair Part
No:
TUB0053
TUB0003
TUB0068
TUB0014
TUB0017
CERTIFICATE
2
Sensitivity
3
0.1 mg/m
3
0.1 mg/m
0.2-6 ppm
1-25 ppm
Sensitivity
dependant
compound present
on