Appendix 2 - Using the test port for other Draeger tubes

As well as the Draeger oil Impactor the F6001 test port can also be used a range of other tubes listed
below:
Test
Oil (chemical reagent tube)
Nitrous Fumes
Sulphur Dioxide
Polytest
To prepare the tubes Factair recommends the F2187 Draeger tube tip cutter.
If you have another Draeger tube, which you wish to use with the instrument, contact Factair for
technical advice on its suitability.
If you have no knowledge of the type of compressor lubricant being used in the breathing-air system
then Factair recommends 15 minutes as this covers the vast majority of lubricants. Once you have
selected the tube to tested follow the testing process as described earlier in the manual.
Draeger ref.
6728371
NO
8103661
x
SO
6728491
2
CH28401
14
Factair Part
Sensitivity
No:
TUB0003
0.1 mg/m
TUB0068
0.2-6 ppm
TUB0014
1-25 ppm
TUB0017
Sensitivity
compound present
To complete a tube test commence a new test and
after the unit had calibrated the oxygen cell and
you have input the test location details you will
have the option to select a tube test as shown,
press the button to proceed.
In this sub menu you will then be able to choose
from oil, sulphur dioxide, nitrous fumes or the
polytest tube.
If you are intending to complete an oil tube test
then you will need to input the test time. If you can
determine with a high degree of certainty the
specific compressor lubricant being used, then
refer to the "test times" chart provided on
Factair's website marked for the F6000.