Advertisement

Oil Impactor

The Oil Impactor's protective seal must be kept in place for the duration of the test and only
removed after the test has been completed.
With a standard measurement range of 0.1 to 1.0 mg/m
3
0.05 mg/m
. The Impactor can detect all mineral and synthetic oil mists, it features a series of 3
horizontal lines, each formed by a series of precision manufactured nozzles. These horizontal lines of
nozzles are calibrated to a different concentration of oil. When air is passed through these nozzles any
oil present in the air is deposited on the glass plate prior to the air being exhausted by vents around its
circumference, this allows the user to easily and quickly identify minute quantities of oil.
For tests that pass with an oil concentration of less than 0.05mg/m
tests above that read as follows:
The bottom line represents an oil concentration of 0.1 mg/m
continuous line then the concentration is in excess of 0.1 mg/m
concentration of 0.5 mg/m
in excess of 0.5 mg/m
continuous line the concentration is in excess of 1.0 mg/m
Pass* -
greater
than
3
0.05mg/m
but less
than
3
0.1mg/m
Fail* -
greater than
3
0.5mg/m
but less than
3
1.0mg/m
* – The above pass and fail criteria is based on breathing-air standard EN12021
Using the Draeger tube tip cutter
3
, again when the oil deposited forms a continuous line the concentration is
3
. The top line represents a concentration of 1.0 mg/m
3
the Impactor has a limit of detection of
3
, the screen will remain blank. For
3
. When the oil deposited forms a
3
. The middle line represents a
3
.
Pass* -
greater
than
3
0.1mg/m
but less
than
3
0.5mg/m
Fail* -
greater
than
3
1.0mg/m
16
3
, again when this forms a