Factair F6001 Operating Manual page 9

Safe-air tester
Table of Contents

Advertisement

Carry out the odour test by depressing the odour
test button and smelling the air exhausted from
the sintered outlet, then release the button. If
there is a serious odour problem, fail the sample
at this point of the test.
Initiating flows of
seriously contaminated air through the tube ports
could result in damage to the unit. If the odour
test is satisfactory, press the 'pass' icon.
The instrument will now commence a 7-minute
purge sequence, during which time an airflow will
be passed through the test port and across the
electronic cells. This clears contaminants from
previous tests and ensures that a representative
sample is being tested.
Once the purge has finished, the Draeger Impactor
should be inserted into the test port and tightened
firmly; ensure the Impactor has an intact
protective seal. This seal must remain in place for
the duration of the test. Press the button in top
right hand corner to start the test.
The test will now commence and the duration will
be displayed as a countdown, together with
pressure, ambient temperature, carbon dioxide,
carbon monoxide and oxygen content.
The unit can be left until the test time has elapsed.
8

Advertisement

Table of Contents
loading