Set ting and Using the Timer
Your LX Immersion Circulator has a timer that can be set for time
periods from 5 minutes to 99 hours. The factory default is 24:00
hours.
1. Press
twice.
3. Press
.
When the timer is running, the remaining time and the set point
temperature appear alternately on the display. When the time
period set has elapsed, the Circulator will beep three times, go
silent for 15-20 seconds and then repeat. The Circulator will
continue circulating the bath fluid and maintaining it at the set
point temperature until the Stop button is pressed; this also
resets the timer.
Loss of Power Restar t
!
If electrical power lost while your Circulator is in use, it will
begin operating automatically once power is restored, with the
temperature displayed in °C and the timer set for 24:00 hours.
14
WARNING: The unit will start automatically after a
disruption in electrical power.
2. Press
and/
or
until the
desired time period
is displayed.