Olympus Vanta L Series Getting Started Manual page 8

X-ray fluorescence analyzer
Hide thumbs Also See for Vanta L Series:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Trademarks
SD, miniSD, and microSD logos are trademarks of SD-3C, LLC.
All brands are trademarks or registered trademarks of their respective owners and third party entities.
Radiation Profiles
The table below represents the upper bounds on the worst case (maximum power and minimum beam
filtration) using a 316 stainless steel target. Note that these beam conditions do not represent typical
use values or a combination of settings typically available from the factory. To convert from µSv/h to
mR/h, divide the value by 10.
Model
conditions
VLW, Element
35 kV,
50 µA,
Al filter
VCR, VCW
40 kV,
100 µA,
open filter
VMR, VMW,
50 kV,
VCA,
80 µA,
Element-S
open filter
Survey updated: January 29 2020
a. BK = background reading (< 1 µSv/h)
EN-6
Beam
Trigger
a
BK
BK
BK
5 cm
10 cm
4 µSv/h
2 µSv/h
25 µSv/h
10 µSv/h
40 µSv/h
25 µSv/h
Vanta — Getting Started Guide
30 cm
BK
BK
2 µSv/h

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents