Chapter 9: Technical Reference

EMC Guidance

Safety Standards: IEC 60601-1, IEC 60601-2-50, EMC Standards IEC 60601-1-2
Table 1
Guidance and manufacturer's declaration - electromagnetic immunity
The BiliSoft Phototherapy System is intended for use in the electromagnetic environment specified below.
The customer or the user of the BiliSoft Phototherapy System should assure that it is used in such an
environment.
Immunity Test
Electrostatic discharge
(ESD) IEC 61000-4-2
Electrical fast transient/
burst IEC 61000-4-4
34
Small Fiberoptic Light Pad
15cm.
Effective Surface Area
Large Fiberoptic Light Pad
25cm.
Effective Surface Area
IEC 60601 Test Level
+/- 6kV contact +/- 8kV
air
+/- 2kV for power supply
lines.
+/- 1kV for input/output
lines.
M1110150 006
30cm.
30cm.
Compliance level
+/- 6kV contact
+/- 8kV air
+/- 2kV for power supply
lines.
+/- 1kV for input/output
lines.
© 2007 by General Electric Company. All rights reserved.
Electromagnetic
environment-guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material the
relative humidity should
be at least 30%.
Mains power quality
should be that of a typical
commercial or hospital or
domestic environment.