Thermo Scientific Barnstead EASYpure RODI 1332 Series Operation Manual And Parts List page 43

Table of Contents

Advertisement

Problem (cont.)
Tank not filling (low pressure indicated by
gauge.)
Low RO purity LED illuminated.
Decrease in % rejection.
Decrease in system productivity.
RO prefilter plugs rapidly.
Possible Causes (cont.)
Feedwater supply valve closed.
RO prefilter plugged.
Solenoid valve not open.
Feed pressure below specification.
New RO membrane rinsing up.
Improperly installed RO membrane.
Fouled RO membrane.
Decrease in water temperature.
RO membrane fouled.
Low incoming water pressure.
Low RO membrane operating
pressure.
High turbidity feedwater.
T
ROUBLESHOOTING
Action (cont.)
Open feedwater supply valve.
Replace RO prefilter.
Call Technical Service.
Increase pressure.
Continue rinse-up.
Be sure RO membrane is installed
properly.
Replace RO membrane.
If decline in flowrate is unacceptable,
install feedwater temperature valve
to elevate water temperature. Valve
available by calling customer service.
Replace RO membrane.
Be sure prefilter is not blocked.
Increase incoming water pressure.
Increase operating pressure. If
needed, use Barnstead accessory
AY1332X1 external pressure boost
assembly.
Call Technical Service. If needed, use
Barnstead accessory AY1332X2
External Filter Assembly with gauges.
This contains high volume 5 micron
particulate filtering in conjunction
with membrane antiscalant.