Turn the unit around to provide access to the
drain plug on the lower edge of the back panel.
d.
Place the drain plug over a bucket or other suit-
able large container. Remove the drain plug by
turning it while pulling until it comes out.
e.
Drain remaining water from the reservoir and
system.
f.
Replace the drain plug, taking care to fully
insert it into the drain fitting and then back out
very slightly. Reconnect power and water.
7.
Put the unit into Run Mode and allow the
reservoir to automatically refill and recirculate
the water through the system.
8.
After the "RO System Operating" LED is no
longer illuminated, drain the system again as
described in step 6 of this section, discarding
the water.
9.
Place unit in "IdL" and remove the empty
cartridges according to the instructions in the
Cartridge Removal section. Drain and retain
the empty cartridge tubes for future use.
Remove the hose barb from the draw-off valve.
Keep the hose barb for future use.
10. Install and rinse new cartridges according to
the instructions in the DI Cartridge
Replacement and Rinse Up section. Do not
reinstall used cartridges or 0.2 micron final fil-
ter (they may contain large amounts of bacte-
ria.)
11. Install and rinse new 0.2 micron final filter
according to the instructions in the 0.2 Micron
Final Filter Replacement section of this man-
ual.