Thermo Scientific Barnstead EASYpure RODI 1332 Series Operation Manual And Parts List page 36

Table of Contents

Advertisement

M
S
AINTENANCE AND
Note
The UV lamp contains mercury. If
broken or no longer needed, do not
dispose of the UV lamp in the trash.
Recycle or dispose of the UV lamp as
hazardous waste.
Note
If UV lamp is not installed properly an
"Er3" message will appear for 15 sec-
onds of every minute during Run
Mode.
36
ERVICING
4.
Remove the screws securing the EASYpure
R
D
top cover.
O
I
5.
Remove the top cover by lifting straight up.
6.
Locate the UV oxidation chamber and pull the
top black plastic cover off. Do not pull on the
cable.
7.
While holding on to the lamp, remove the plug
from the lamp. Dispose of the UV lamp appro-
priately. The used lamp does contain mercury
vapor and should not be disposed of in the
trash. Recycle or dispose of the used lamp as
hazardous waste.
8.
Remove the replacement lamp from its con-
tainer. DO NOT TOUCH THE GLASS POR-
TION OF THE LAMP. It is recommended that
lint free gloves be worn when handling the
lamp. The glass portion must be free of finger-
prints, perspiration, etc. Even a light coating of
perspiration will reduce the effectiveness of the
lamp.
9.
Clean the lamp with isopropyl alcohol and a lint
free cloth.
10. Carefully insert and hold the UV lamp partially
into the UV chamber.
11. Connect the UV lamp to the receptacle in the
black plastic cover of the UV chamber.
Replace the black plastic cover on the UV
chamber.
12. Reinstall the EASYpure R
latch cartridge access door closed.
13. Reinstall the Ventgard cap.
14. Reattach the power cord and reconnect the
unit to the power supply.
D
top cover and
O
I