COMBITEST422 TABLE OF CONTENTS PRECAUTIONS AND SAFETY MEASURES ............... 4 1.1. Preliminary instructions ....................4 1.2. During use ........................5 1.3. After use .......................... 5 1.4. Definition of measurement (overvoltage) category ............5 GENERAL DESCRIPTION ................... 6 2.1. Instrument functions ......................6 PREPARATION FOR USE ...................
Page 3
COMBITEST422 6.7.10. Verification of protection against indirect contacts (TN systems) .......... 69 6.7.11. Anomalous situations ......................71 6.8. LoZ: Line/Loop impedance with high resolution ............. 75 6.9. 1,2,3: Phase sequence and phase concordance test ............. 76 6.9.1. Anomalous situations ......................79 6.10.
COMBITEST422 1. PRECAUTIONS AND SAFETY MEASURES The instrument has been designed in compliance with guidelines IEC/EN61557, BS7671 17th and 18th editions and IEC/EN61010, relevant to electronic measuring instruments. Before and after carrying out the measurements, carefully observe the following instructions: •...
COMBITEST422 1.2. DURING USE Please carefully read the following recommendations and instructions: CAUTION Failure to comply with the caution notes and/or instructions may damage the instrument and/or its components or be a source of danger for the operator. • Before changing function, disconnect the test leads from the circuit under test.
COMBITEST422 2. GENERAL DESCRIPTION 2.1. INSTRUMENT FUNCTIONS The instrument can perform the following tests: • RPE Continuity test of earth, protective and equipotential conductors with test current higher than 200mA and open-circuit voltage between 4V and 24V. • M Kommenterede [S2]: dato che, più avanti nella frase, si Measurement of insulation resistance with continuous test voltage of specifica DC, “continuous”...
COMBITEST422 3. PREPARATION FOR USE 3.1. INITIAL CHECKS Before shipping, the instrument has been checked from an electric as well as a mechanical point of view. All possible precautions have been taken so that the instrument is delivered undamaged. However, we recommend checking it to detect any damage possibly suffered during transport.
COMBITEST422 ON/OFF key to switch on/off the instrument ESC key to exit the selected menu without confirming MENU key to go back to the general menu at any time keys to move the cursor through the different screens in order to...
COMBITEST422 5. GENERAL MENU Pressing the MENU/ESC key in any condition of the instrument allows going back to the general menu in which internal parameters may be set and the desired measuring function may be selected. 15/10 – 18:04 15/10 – 18:04...
COMBITEST422 5.1.2. Country Move the cursor to Country by means of the arrow keys 15/10 – 18:04 S E T (,) and confirm with ENTER. Subsequently, the display E u r o p e shows the screen which allows selecting the country of...
COMBITEST422 keys (,) and confirm with ENTER. Subsequently, the Auto Power Off : display shows the screen which allows enabling/disabling the : OFF Keys Beep auto power off function, the sound of the function keys and : ...
COMBITEST422 6. OPERATING INSTRUCTIONS 6.1. AUTO: AUTOMATIC TEST SEQUENCE (RA , RCD, M) This function allows performing the following measurements in an automatic sequence: ➢ Overall earth resistance without causing the RCD’s tripping (Ra ) ➢ Tripping current and tripping time of General RCDs type A (...
Page 14
COMBITEST422 TN systems Press the MENU key, move the cursor to AUTO in the 15/10 – 18:04 A U T O > < main menu by means of the arrow keys (,) and confirm with ENTER. Subsequently the instrument Z L - N = - - - Ω...
Page 15
COMBITEST422 Note the correct voltage values between L-N and L-PE as 15/10 – 18:04 A U T O > < shown in the screen to the side. Z L - N = - - - Ω I s c = - - - A Z L - P E = - - - Ω...
Page 16
COMBITEST422 The insulation test starts and the screen to the side 15/10 – 18:04 A U T O > < appears on the display. The RL-N, RL-PE and RN-PE values appear on the display. > 9 9 9 M ...
Page 17
COMBITEST422 TT/IT systems 1. Press the MENU key, move the cursor to AUTO in the 15/10 – 18:04 A U T O > < main menu by means of the arrow keys (,) and confirm with ENTER. Subsequently, the instrument R A = - - - ...
Page 18
COMBITEST422 probe. The instrument will start the automatic test sequence. CAUTION If message “Measuring…” appears on the display, the instrument is performing measurement. During this whole stage, do not disconnect the test leads of the instrument from the mains. 15/10 – 18:04...
Page 19
COMBITEST422 In case of a negative result of the Ra test (see § 12.8), 15/10 – 18:04 A U T O > < the auto test is automatically blocked, the message “NOT OK” is shown and the screen to the side is displayed.
COMBITEST422 6.1.1. Anomalous situations 1. If the instrument detects a L-N or L-PE voltage higher 15/10 – 18:04 A U T O than the maximum limit (265V), it does not carry out the Z L - N = - - - Ω...
COMBITEST422 6.2. DMM: DIGITAL MULTIMETER FUNCTION This function allows reading the real time TRMS values of P-N Voltage, P-PE Voltage, N- PE Voltage and Frequency (@ P-N inputs) when the instrument is connected to an installation. Fig. 6: Instrument connection through mains plug Fig.
Page 22
COMBITEST422 3. The TRMS values of L-N voltage, L-PE voltage, N-PE 15/10 – 18:04 D M M voltage and the frequency of L-N voltage are shown on the display. FREQ. = 50.00 Press the GO/STOP key to enable/disable the “HOLD”...
COMBITEST422 6.3. RPE: CONTINUITY OF PROTECTIVE CONDUCTORS This function is performed in compliance with standards IEC/EN61557-4, BS7671 17th/18th edition and allows measuring the resistance of protective and equipotential conductors. CAUTION • The instrument can be used for measurements on installations with overvoltage category CAT IV 300V to earth and max 415V between inputs.
Page 24
COMBITEST422 Fig. 9: Continuity test by means of remote switch probe Press the MENU key, move the cursor to RPE in the main 15/10 – 18:04 R P E menu by means of the arrow keys (,) and confirm with - - - ...
COMBITEST422 CAUTION Always make sure, before any test, that the compensation resistance value of the cables is referred to the cables currently used. In case of doubt, repeat the cable calibration procedure as indicated in § 6.3.2. Press the GO/STOP key on the instrument or the START key on the remote switch probe.
COMBITEST422 At the end of the set duration time, the instrument shows 15/10 – 18:04 R P E on the display the maximum value among all the partial 0 . 5 4 measurements performed and the message "OK" in case...
COMBITEST422 6.3.3. Anomalous situations In case the detected value is higher than the set limit, the 15/10 – 18:04 R P E instrument gives a long acoustic signal and displays a 4 . 5 4 screen similar to the one reported here to the side.
COMBITEST422 6.4. LO: CONTINUITY OF PROTECTIVE CONDUCTORS WITH 10A This function allows measuring the resistance of protective and equipotential conductors with a test current >10A by using the Lo 10A optional accessory EQUITEST connected to the instrument through the C2050 cable. The optional accessory EQUITEST must be directly powered by the mains on which measurements are performed.
COMBITEST422 to the side is shown on the display indicating the E Q U I T E S T S N : 2 0 0 9 0 0 1 1 information relevant to the accessory. F W : 1 . 0 0 H W : 1 .
COMBITEST422 terminals, it does not perform the test, it emits a prolonged - - - acoustic signal and displays a screen like the one to the side. I t e s t - - - A V i n > 3 V 0.500 ...
Page 31
COMBITEST422 conductors and between each active conductor and the earth. The following operating modes are available: • MAN the test can be carried out between the L-N, L-PE or N-PE conductors and has a fixed duration of 3s when the GO/STOP key is pressed on the instrument (or START on the remote switch probe).
Page 32
COMBITEST422 Fig. 13: Insulation between L-N-PE by means of mains plug (MAN and AUTO) Fig. 14: Insulation between L-PE by means of mains plug (TMR mode) Fig. 15: Insulation between L-PE by means of single cables (TMR mode) EN - 32...
Page 33
COMBITEST422 Fig. 16: Insulation between L-PE by means of single cables and remote switch probe (TMR mode)) M 15/10 – 18:04 Press the MENU key, move the cursor to M in the main menu by means of the arrow keys (,) and confirm with - - - M ...
COMBITEST422 • Disconnect any cable not strictly involved in measurement. • Before connecting the test leads, make sure that there is no voltage at the ends of the conductors to be tested. 5. Connect the test leads and remote switch probe to the ends of the conductors to be tested as shown in Fig.
COMBITEST422 "Mode" section. The instrument displays a screen like the - - - M one shown to the side. Set the duration of the Vt = - - - V T = - - - s measurement in the "Time" section and follow the steps from point 2 to point 5 of §...
COMBITEST422 the "Mode" section. The instrument displays a screen like - - - M Vt = - - - V RL-N the one shown to the side. Set the duration of the RL-PE - - - M Vt = - - - V measurement in the "Time"...
Page 37
COMBITEST422 emits a long acoustic signal to indicate the negative result 0 . 0 1 M of the test and displays a screen like the one at the side. NOT OK 500V 1.00M L-PE MODE Vtest Lim. FUNC M ...
Page 38
COMBITEST422 terminals, it does not perform the test, emits a prolonged - - - M acoustic signal and displays a screen like the one at the side. - - - V - - - s Vin >10V 500V 1.00M...
COMBITEST422 6.6. RCD: TEST ON DIFFERENTIAL SWITCHES This function is performed in compliance with standard IEC/EN61557-6, BS7671 17th/18th edition and allows measuring the tripping time and current of molded-case differential switches of type A ( ) , AC ( ) or B ( ) being General (G) and Selective (S).
Page 40
COMBITEST422 Fig. 17: Connection for single-phase 230V system by means of mains plug Fig. 18: Connection for single-phase 230V system with single cables and remote switch probe Fig. 19: Connection for 400V + N + PE three-phase system by means of single cables and...
Page 41
COMBITEST422 Fig. 20: Connection for a 400V + N (no PE) three-phase system by means of single cables and remote switch probe Fig. 21: Connection for a 400V + PE (no N) system with cables and remote switch probe Press the MENU key, move the cursor to RCD in the 15/10 –...
COMBITEST422 3. Insert the green, blue and black connectors of the three-pin plug cable into the corresponding inputs B3, B4 and B1 of the instrument. As an alternative, use the single cables and apply the relevant alligator clips to the free ends of the cables. It is also possible to use the remote switch probe by inserting its multipolar connector into the input B1.
COMBITEST422 7. In case of a positive result (all tripping times comply 15/10 – 18:04 R C D with what indicated in 12.4) of all the test sequentially 0 ° 1 8 0 ° performed, the “OK” message is shown and the screen...
COMBITEST422 7. In case of a positive result (all tripping times comply 15/10 – 18:04 R C D with what indicated in 12.4) of all the tests sequentially 0 ° 1 8 0 ° performed, the “OK” message is shown and the screen to the side is displayed by the instrument.
COMBITEST422 6.6.4. mode The standard defines the tripping times for RCDs at nominal current. The mode is used to detect the tripping time at tripping current (which could also be lower than the nominal current). 15/10 – 18:04 Press the GO/STOP key on the instrument, the START...
COMBITEST422 6.6.5. Anomalous situations If the instrument detects a frequency higher than the 15/10 – 18:04 R C D maximum limit (63Hz), it does not carry out the test and - - - displays a screen like the one to the side.
Page 47
COMBITEST422 If the instrument detects that the phase L and neutral N 15/10 – 18:04 R C D leads are inverted, it does not carry out the test and a - - - screen similar to the one reported to the side is displayed.
Page 48
COMBITEST422 If the instrument detects the absence of the signal to 15/10 – 18:04 R C D terminal B1 (phase conductor), it provides the warning - - - screen shown to the side and blocks the execution of the tests.
COMBITEST422 6.7. LOOP: LINE/LOOP IMPEDANCE AND OVERALL EARTH RESISTANCE This function is performed in compliance with standard IEC/EN61557-3, BS7671 17th/18th edition and allows measuring the line impedance, the fault loop impedance and the prospective short-circuit current. CAUTION Depending on the selected electrical system (TT, TN or IT), some connections and function modes are disabled by the instruments (see Table 1: Conditions of positive outcome depending on the test parameters).
Page 50
COMBITEST422 Fig. 23: P-N/P-PE test for single-phase/two-phase systems with cables and remote probe Fig. 24: P-N/P-PE test for 400V+N+PE three-phase with single cables and remote probe Fig. 25: P-P measurement for 400V+N+PE three-phase systems Fig. 26: P-PE/P-N test for 400V + PE systems by means of single cables and remote probe...
Page 51
COMBITEST422 Fig. 27: P-PE measurement for IT systems by means of single cables and remote probe Fig. 28: P-PE 2-wire test for single-phase/two-phase 230V systems with mains plug Fig. 29: P-PE 2-wire test for single-phase/two-phase systems with cables and remote probe Fig.
COMBITEST422 6.7.1. Test types The protection of electrical lines is the essential part of a project to guarantee correct functionality and avoid injury to persons or damage of property. To this purpose, the safety guidelines impose on electrical designers to also design the electrical installation in order to obtain: 1.
Page 53
COMBITEST422 Mode Condition x OK outcome Condition x OK outcome Condition x OK outcome No outcome No outcome No outcome Br.Cap Isc L-L max < BC Isc L-L max < BC Isc L-L max < BC (IscL-Lmin 2F) →Tmax → Tmax <...
COMBITEST422 6.7.2. Test cable calibration (ZEROLOOP) In order to obtain better results, it is strongly recommended to perform the preliminary calibration of the test cables or of the cable with mains plug by using the ZEROLOOP accessory before performing the test. In this way, the instrument automatically subtracts the resistance of the test cables, providing the actual result on the display.
Page 55
COMBITEST422 The value of the test leads/mains plug resistance is 15/10 – 18:04 L O O P maintained by the instrument up to the reset operation - - - performed by the user (for example, for the insertion of ...
COMBITEST422 6.7.3. STD Mode – Generic test This mode performs the impedance measurement and the calculation of prospective short- circuit current without any evaluation. Therefore, at the end of the test, no outcome is given by the instrument. 15/10 – 18:04...
Page 57
COMBITEST422 Press the GO/STOP key on the instrument, the START 15/10 – 18:04 L O O P key on the remote switch probe or use the AutoStart Ipfc - - - feature (see § 5.1.5). The instrument starts the measurement and the “Measuring…” message is shown ...
COMBITEST422 6.7.4. Br.Cap mode – Verification of the breaking capacity of protection devices Press the MENU key, move the cursor to LOOP in the 15/10 – 18:04 L O O P main menu by means of the arrow keys (,) and - - - confirm with ENTER.
Page 59
COMBITEST422 Press the GO/STOP key on the instrument, the START 15/10 – 18:04 L O O P key on the remote switch probe or use the AutoStart - - - feature (see § 5.1.5). The instrument starts the measurement and the “Measuring…” message is shown ...
COMBITEST422 6.7.5. TripT - Verification of protection coordination Press the MENU key, move the cursor to LOOP in the 15/10 – 18:04 L O O P main menu by means of the arrow keys (,) and - - - confirm with ENTER. Subsequently, the instrument displays a screen similar to the one reported here to the ...
Page 61
COMBITEST422 Note the presence of the correct voltage values between 15/10 – 18:04 L O O P L-L and L-PE corresponding to the selections carried out - - - in the initial phase (see § 5.1.3) as shown in the screen to the side.
COMBITEST422 6.7.6. Ra 2-wire test - Verification of protection against indirect contacts Press the MENU key, move the cursor to LOOP in the 15/10 – 18:04 L O O P main menu by means of the arrow keys (,) and - - - confirm with ENTER.
Page 63
COMBITEST422 Note the presence of the correct voltage values between 15/10 – 18:04 L O O P L-PE corresponding to the selections carried out in the - - - initial phase (see § 5.1.3) as shown in the screen to the side.
COMBITEST422 6.7.7. Ra 3-wire test - Verification of protection against indirect contacts 1. Press the MENU key, move the cursor to AUTO in the 15/10 – 18:04 L O O P main menu by means of the arrow keys (,) and confirm with ENTER.
Page 65
COMBITEST422 shown in the screen to the side. Z L - N = - - - Ω I s c = - - - A Z L - P E = - - - Ω I f c = - - - A F R E Q = 5 0 .
COMBITEST422 6.7.8. Verification of protection against indirect contacts (IT systems) Press the MENU key, move the cursor to LOOP in the 15/10 – 18:04 L O O P main menu by means of the arrow keys (,) and Ipfc - - - confirm with ENTER.
COMBITEST422 <50V or <25V), the “OK” message and the screen to the Ipfc side are displayed by the instrument. The screen contains the value of the first fault current measured, expressed in mA (see § 12.9). FREQ. = 50.00Hz VL-N=234V...
Page 68
COMBITEST422 main menu by means of the arrow keys (,) and - - - confirm with ENTER. Subsequently, the instrument displays a screen similar to the one reported here to the - - - side. FREQ. = 0.00Hz Select “Europe” as a country (see § 5.1.2), the options VL-PE=0V “TT”, “25 or 50V”, “50Hz or 60Hz”...
COMBITEST422 key on the remote switch probe or use the AutoStart - - - feature (see § 5.1.5). The instrument starts the measurement and the “Measuring…” message is shown - - - on the display. During this whole stage, do not disconnect the measuring leads of the instrument from the system FREQ.
Page 70
COMBITEST422 main menu by means of the arrow keys (,) and - - - confirm with ENTER. Subsequently, the instrument displays a screen similar to the one reported here to the - - - ZL-PE side.Select “Europe” as a country (see § 5.1.2), the options “TN”, “25 or 50V”, “50Hz or 60Hz”...
COMBITEST422 L-N and L-PE corresponding to the selections carried out - - - in the initial phase (see § 5.1.3) as shown in the screen to the side. - - - ZL-PE FREQ. = 50.00Hz VL-N=231V VL-PE=232V L-PE 0.2s...
Page 72
COMBITEST422 maximum limit (63Hz), it does not carry out the test and Ipfc - - - displays a screen like the one to the side. ZL-PE - - - FREQ. = >63Hz VL-N=0V VL-PE=0V Freq. out of range L-PE...
Page 73
COMBITEST422 conductor, it provides the warning screen shown to the Ipfc - - - side and blocks the execution of the tests. Check the PE conductor and earth connection efficiency. ZL-PE - - - FREQ. = 50.00Hz VL-N= 234V...
Page 74
COMBITEST422 leads are inverted, it does not carry out the test and a Ipfc - - - screen similar to the one reported to the side is displayed. Rotate the mains plug or check the connection of ZL-PE - - - measuring cables.
COMBITEST422 6.8. LOZ: LINE/LOOP IMPEDANCE WITH HIGH RESOLUTION This Line/loop impedance measurements with high resolution (0.1m) is performed by using the optional accessory IMP57 connected to the Master unit through the optical cable/RS-232 C2001 supplied with same accessory. The IMP57 must be directly powered by the mains on which measurements are performed.
COMBITEST422 6.9. 1,2,3: PHASE SEQUENCE AND PHASE CONCORDANCE TEST This function allows testing the phase sequence and concordance with 1-wire method by direct contact with live parts (not on cables with insulating sheath). Fig. 31: Phase sequence check with test leads Fig.
Page 77
COMBITEST422 Press the GO/STOP key on the instrument or the START 15/10 – 18:04 1 2 3 key on the remote switch probe. The instrument will start the test. - - - The "Touch L1" message is shown on the display to...
Page 78
COMBITEST422 At the end of the test, if the detected phase sequence is 15/10 – 18:04 1 2 3 incorrect, the instrument displays a screen like the one shown to the side (result "213") and the “NOT OK” message. 2 1 3...
COMBITEST422 6.9.1. Anomalous situations If the instrument detects an input voltage frequency 15/10 – 18:04 1 2 3 exceeding the allowed full scale, it will display a screen like the one to the side. - - - Freq. out of range...
COMBITEST422 6.10. V%: VOLTAGE DROP OF MAINS Kommenterede [S20]: forse MAINS nel senso di rete elettrica? This feature allows evaluating the percentage value of voltage drop between two points of mains in which a protection device is installed and comparing this value to possible limit Kommenterede [S21]: idem come commentato sopra? values specified by guidelines.
Page 81
COMBITEST422 Press the MENU key, move the cursor to V% in the V % 15/10 – 18:04 main menu by means of the arrow keys (,) and V% - - - confirm with ENTER. Subsequently, the instrument displays a screen similar to the one reported here to the ...
Page 82
COMBITEST422 Use the , keys and move the cursor to the “Z><” V % 15/10 – 18:04 position. Press the GO/STOP key on the instrument to V% - - - start the test. The result of the Z1(offset ) measurement is shown on the display above the “Z><”...
COMBITEST422 6.10.1. Anomalous situations V % If the instrument detects a frequency higher than the 15/10 – 18:04 maximum limit (63Hz), it does not carry out the test and V% - - - displays a screen like the one to the side.
Page 84
COMBITEST422 V % If the instrument detects a dangerous voltage on PE 15/10 – 18:04 conductor, it provides the warning screen shown to the V% - - - side and blocks the execution of the tests. Check the PE conductor and earth connection efficiency.
Page 85
COMBITEST422 V % If the instrument detects that the phase L and neutral N 15/10 – 18:04 leads are inverted, it does not carry out the test and a V% - - - screen similar to the one reported to the side is displayed.
COMBITEST422 7. STORING RESULTS The instrument allows saving max 999 measured values. The saved data can be recalled to the display and deleted at any moment, and, upon saving, they can be associated with up to a maximum of 3 levels of numeric markers relevant to the installation name, the PV string and the PV module (with max value 250).
COMBITEST422 7.2. RECALL OF DATA TO DISPLAY AND MEMORY DELETION 15/10 – 18:04 1. Position the cursor onto MEM by using the arrow keys D a t e T y p e (,) and confirm with ENTER. The screen to the side...
COMBITEST422 8. CONNECTING THE INSTRUMENT TO THE PC CAUTION • The connection between instrument and PC is realized by means of cable C2006. • In order to transfer the data onto a PC, it is necessary to install beforehand both the management software and the drivers of cable C2006 on the PC itself.
COMBITEST422 9. MAINTENANCE 9.1. GENERAL INFORMATION ➢ While using and storing the instrument, carefully observe the recommendations listed in this manual in order to prevent possible damage or danger during use. ➢ Do not use the instrument in environments with high humidity levels or high temperatures.
COMBITEST422 10. TECHNICAL SPECIFICATIONS Accuracy is calculated as: ±[%reading + (no. of digits) * resolution] at 23°C, <80%RH 10.1. TECHNICAL CHARACTERISTICS AC TRMS voltage Range [V] Resolution [V] Accuracy 15 460 (3%rdg + 2dgt) Frequency Range [Hz] Resolution [Hz] Accuracy 47.50 ...
Page 91
COMBITEST422 Differential protection type (RCD): AC ( ), A ( ),General (G), Selective (S) and B( 100V 265V RCD type A, A and B (IN ≤100mA) Voltage range P-PE, P-N: 190V 265V RCD type B (IN = 300mA) Voltage range N-PE : <10V...
COMBITEST422 100V 265V Voltage range P-PE, P-N: Voltage range N-PE: <10V 50/60Hz 5% Frequency: Overall earth resistance in systems with Neutral (3-wire) – (30mA or higher RCD) Accuracy Range [] Resolution [] 0.05 9.99 0.01 (5rdg +8dgt) 10.0 ...
COMBITEST422 11. SERVICE 11.1. WARRANTY CONDITIONS This instrument is warranted against any material or manufacturing defect, in compliance with the general sales conditions. During the warranty period, defective parts may be replaced. However, the manufacturer reserves the right to repair or replace the product.
COMBITEST422 12. THEORETICAL APPENDIXES 12.1. CONTINUITY OF PROTECTIVE CONDUCTORS Check the continuity of: • Protective conductors (PE), main equalizing potential conductors (EQP), secondary equalizing potential conductors (EQS) in TT and TN-S systems • Neutral conductors having functions of protective conductors (PEN) in TN-C systems.
COMBITEST422 Check that the insulation resistance of the installation complies with the requirements of the applicable guidelines. This test has to be performed with the circuit being tested not powered and with the possible loads it supplies disconnected. Allowable values The values of the measured voltage and of the minimum insulation resistance can be taken from the following table.
COMBITEST422 polarization. Because of the polarized molecules, a polarization (absorption) current generates, which lowers the total value of insulation resistance. Parameter PI consists in the ratio between the value of insulation resistance measured after 1 minute and after 10 minutes. The test voltage is maintained throughout the whole...
Page 98
COMBITEST422 (double or reinforced insulation or earthed metal screen) and absence of earthed points (insulated from the earth). A PELV system is a zero-category system or protective extra low voltage system characterized by power supply from an independent (e.g. batteries, small generator set) or safety source (e.g.
Page 99
COMBITEST422 EXAMPLE OF SEPARATION TEST BETWEEN ELECTRICAL CIRCUITS Insulation or safety transformer separating the circuits TEST BETWEEN ACTIVE PARTS Connect an instrument probe to one of the two conductors of the separated circuit and the other to one of the conductors of a non- separated circuit.
COMBITEST422 12.4. TEST ON DIFFERENTIAL SWITCHES (RCD) Purpose of the test Checking that the General (G) and Selective (S) differential protection devices have been correctly installed and adjusted and that they maintain their characteristics over time. The check must make sure that the differential switch trips at a current not higher than its...
COMBITEST422 Measurement of tripping current for protection differential switches ➢ This test aims at checking the real tripping current of general differential switches (it does not apply to selective differential switches). ➢ In the presence of differential switches with selectable tripping current, it is useful to perform this test in order to check the real tripping current of the differential switch.
COMBITEST422 12.6. VERIFY OF PROTECTION AGAINST INDIRECT CONTACTS IN TN SYSTEMS Purpose of the test Protection against indirect contacts in TN systems must be guaranteed by means of a protection device against overcurrents (typically MCB or fuse), which switches off the...
Page 103
COMBITEST422 Depending on the set values of phase-phase, phase-neutral or phase-PE voltage (see § 5.1.3) and the measured value of fault loop impedance, the instrument calculates the minimum value of the prospective short-circuit current to be interrupted by the protection device.
COMBITEST422 12.7. RA TEST IN TN SYSTEMS Protection against indirect contacts in TN systems must be guaranteed by means of a protection device against overcurrents (typically MCB or fuse) which switches off the power supply of the circuit or the electrical equipment in case of faults between an active part and a ground mass or a protection conductor within a interval not exceeding 5s, sufficient for the equipments.
COMBITEST422 12.8. VERIFY OF PROTECTION AGAINST INDIRECT CONTACTS IN TT SYSTEMS Purpose of the test Checking that the protection device is coordinated with the value of earth resistance. We cannot assume a priori a reference limit value for earth resistance when checking the measurement’s result.
COMBITEST422 12.9. VERIFY OF PROTECTION AGAINST INDIRECT CONTACTS IN IT SYSTEMS In IT systems, the active parts must be isolated from the ground or be connected to earth through an impedance of sufficiently high value. In case of a single earth fault, the first fault current is weak, and therefore it is not necessary to interrupt the circuit.
COMBITEST422 12.10. VERIFY OF PROTECTION COORDINATION L-L, L-N AND L-PE Purpose of the test Testing the coordination of protective devices (typically MCB or fuse) present in a single- phase or three-phase installation as a function of the limit time of fault extinction by the protection set by the user and the calculated value of the short-circuit current.
COMBITEST422 CAUTION The instrument must be used to measure fault loop impedance values at least 10 times higher than the resolution value of the instrument in order to minimize errors. Depending on the set values of nominal voltage (see § 5.1.3) and the measured value of fault loop impedance, the instrument calculates the minimum value of the prospective short-circuit current to be interrupted by the protection device.
Page 109
COMBITEST422 ➢ Verify the capability of an existing mains line to supply a load ➢ Dimension a new installation ➢ Search for possible causes of troubles on devices, loads, etc. connected to mains Purpose of the test Measure the maximum percentage value of voltage drop between two points of mains.
Need help?
Do you have a question about the COMBITEST422 and is the answer not in the manual?
Questions and answers