Step 2: After selecting the Therapy Setting Mode, press the
right side and select from APAP mode into CPAP mode.
Step 3: After selecting the CPAP Mode, rotate the
in 4.0 cmH
O.
2
Note: Remember to set your ramp time at 0 to save your test time. Please use the same order to
check the air pressure test point 10.0 cmH
adjusting the pressure, please press the
to the right is to increase the pressure, and the
pressure.
Step 4: Press the
± 0.5 cmH
O; please do the pressure calibration again if the pressure value is out of the range.
2
button to the first pressure test point and check if the pressure value accuracy is in
button to select mode. Rotate the
button to the right to choose therapy pressure
O and air pressure test point 20.0 cmH
2
button to select the pressure, the
button to the left is to decrease the
O; while
2
button