Step 7.
Third Point
Step 8.
Adjust
Step 9.
F18
Step 10. The End
After finishing the second calibration, press the
button into the third calibration point
<The picture is for reference only>
Adjust the pressure value by rotating the
until the pressure value on pressure meter shows
20.0 cmH
O.
2
The pressure error value should be in
20.0 ± 0.1 cmH
O.
2
To make the sensor record the pressure completely,
the flow should be stable for about
<The picture is for reference only>
After finishing the third calibration point, press
the
button. The screen will display 18.0.
<The picture is for reference only>
Press the
button again. The screen will display
"Calibration Save", which means calibration
process, is completely.
<The picture is for reference only>
20.0 cmH
O
2
button
10
seconds.