Maintenance; Maintenance Schedule; Disinfecting And Cleaning The Patient Restraints; Disinfecting The Roller System - Ferno 156 User Manual

Roller system
Table of Contents

Advertisement

6.1 Maintenance Schedule

The roller system requires regular maintenance. Set up and follow a
maintenance schedule. The table at right represents minimum intervals
for maintenance.
Keep maintenance records. For a sample record sheet, see "Maintenance
Record" on page 15.
When using maintenance products, follow the manufacturers' directions
and read the manufacturers' material safety data sheets.
6.2 Disinfecting and Cleaning
the Patient Restraints
To remove the patient restraints from the roller system (Figure 13):
1.
To remove the restraints from the roller system, use a small flat-
blade screwdriver to pry off the buckle cover.
2.
Unthread the restraint from the buckle and roller system.
3.
Place the restraints in a mesh laundry bag. Machine wash them
on cold/cold using the delicate cycle and a mild detergent. Allow
them to air dry.
4.
Thread the restraints through the buckle as shown in Figure 13.
Insert the loose end of the strap through the slot in the roller
system (1) and around the edge of the roller-system frame (2).
Orient the buckle as shown. Thread the loose end of the strap
over the buckle roller (3) and under the buckle tab (4).
Orient the buckle cover as shown and attach to the buckle.

6.3 Disinfecting the Roller System

Wipe all surfaces with disinfectant. Follow the disinfectant manufacturer's
instructions for application method and contact time. Ferno recommends
you inspect the roller system for damage as you disinfect it.

6.4 Cleaning the Roller System

1.
Remove the restraints.
2.
Hand clean all surfaces of the roller system with warm water and
a mild detergent.
3.
Rinse with warm, clear water. Dry the roller system with a towel or
allow it to air-dry.
12

Maintenance

6 - MAINTENANCE
Improper maintenance can cause injury. Maintain the
roller system only as described in this manual.
Disinfecting (this page)
Cleaning (this page)
Inspecting (page 13)
Buckle Cover
Roller
Tab
Disinfectants and cleaners containing bleach, phenolics, or
iodines can cause damage. Do not use products containing
these chemicals.
WARNING
Minimum
Maintenance
Each Use
3
4
2
Strap
NOTICE
© Ferno-Washington, Inc. / 234-1845-01 / June 2019
As Needed
Monthly
Buckle
1
13

Advertisement

Table of Contents
loading