Maintenance; Maintenance Schedule; Disinfecting The Fastening System; Cleaning The Fastening System - Ferno iNSLINE Fastening System User Manual

Table of Contents

Advertisement

6.1 Maintenance Schedule

The fastening system requires regular maintenance. Set up and
follow a maintenance schedule. The table at right represents
minimum intervals for maintenance.
Keep maintenance records. For a sample record sheet, see
"Maintenance Record" on page 19.
When using maintenance products, follow the manufacturers'
directions and read the manufacturers' material safety data sheets.

6.2 Disinfecting the Fastening System

Wipe all surfaces with disinfectant. Follow the disinfectant
manufacturer's instructions for application method and contact
time. Ferno recommends you inspect the fastening system for
damage as you disinfect.

6.3 Cleaning the Fastening System

1.
The upper fastener is removable from the floor for cleaning.
See "Removing the Fastening System from the Floor for
Cleaning" on page 14.
2.
Hand clean all surfaces of the fastening system with warm
water and a mild detergent.
3.
Rinse with warm, clear water. Dry the fastening system with a
towel or allow it to air-dry.
4.
Reattach the upper fastening system. See "Attaching the
Fastening System to the Ambulance Floor" on page 15.

6.4 Inspecting the Fastening System

Have your service's equipment maintenance personnel inspect the
fastening system regularly.
Follow the checklist at right and operate the fastening system
through all its functions as described in this manual.
If inspection shows damage or excessive wear, remove the fastening
system from service until repair is made. See "Parts and Service,
Accessories" on page 17.
12
iN∫LINE®

6 - MAINTENANCE

Improper maintenance can cause injury. Maintain the
fastening system only as described in this manual.
Disinfecting (this page)
Cleaning (this page)
Inspecting (this page)
Disinfectants and cleaners containing bleach, phenolics, or
iodines can cause damage. Do not use products containing
these chemicals.
Using abrasive cleaning compounds or applicators can
cause damage. Do not use abrasive materials to clean the
fastening system.
Make sure all connections are tight, but do not overtighten
the hardware. Overtightening, especially with powered
tools, can cause the nut and bolt beneath the ambulance
floor to pull through the support structure.
If inspection indicates the components are pulling through
the support structure, take the fastening system out of
service immediately and install a new fastening system.
WARNING
Minimum
Maintenance
Each Use
NOTICE
Inspection Checklist
Are all components present?
Is the fastening system free of excessive wear?
Are all parts in good condition (no cracks, corrosion, or
damage)?
Do all moving parts operate smoothly and properly?
Does the fastening system lock properly to the floor?
Does the cot load and unload properly?
Does the cot lock properly into the fastening system?
Monitor the ambulance floor, including beneath the
ambulance, and check the integrity of the fastening
system. Is the system worn? Has exposure to the
environment caused noticeable damage?
Is all hardware tight? (Reminder: Do not overtighten
the hardware.)
© Ferno / 234-3621-02 / October 2022
As Needed Monthly

Advertisement

Table of Contents
loading

Table of Contents