Transmitter operation
March 2021
Postrequisites
To clear A010, A011, A012, A013 alerts, you must correct the problem, acknowledge the alert, then
repeat the calibration.
To clear A001, A002, A029, A031 alerts, you must correct the problem, acknowledge the alert, then
power-cycle the transmitter.
For all other alerts:
If the alert is inactive when it is acknowledged, it will be removed from the list.
If the alert is active when it is acknowledged, it will be removed from the list when the alert condition
clears.
12.3.2 View and acknowledge alerts using ProLink III
You can view a list containing all alerts that are active, or inactive but unacknowledged. From this list, you can
acknowledge individual alerts or choose to acknowledge all alerts at once.
Procedure
1. View alerts on the ProLink III main screen under Alerts. If the alerts are not displayed, choose Device
Tools → Alerts.
All active or unacknowledged alerts are listed, and displayed according to the following categories:
Category
Failed: Fix Now
Maintenance: Fix Soon
Advisory: Informational
Notes
All fault alerts are displayed in the Failed: Fix Now category.
All information alerts are displayed in either the Maintenance: Fix Soon category or the Advisory:
Informational category. The category assignment is hard-coded.
The transmitter automatically filters out alerts with Alert Severity set to Ignore.
2. To acknowledge a single alert, check the Ack checkbox for that alert. To acknowledge all alerts at once,
click Ack All.
Postrequisites
To clear A010, A011, A012, A013 alerts, you must correct the problem, acknowledge the alert, then
repeat the calibration.
To clear A001, A002, A029, A031 alerts, you must correct the problem, acknowledge the alert, then
power-cycle the transmitter.
For all other alerts:
If the alert is inactive when it is acknowledged, it will be removed from the list.
108
Description
A meter failure has occurred and must be addressed immediately.
A condition has occurred that can be fixed at a later time.
A condition has occurred, but requires no maintenance from you.