SECTION 7: MAINTENANCE
THERE ARE NO USER-SERVICEABLE PARTS INSIDE THE CAMERA. DAMAGE CAUSED BY
UNAUTHORISED MAINTENANCE OR PROCEDURES WILL INVALIDATE THE WARRANTY.
7.1 R
c
egUlaR
Hecks
The state of the product should be checked regularly, especially the integrity of the PSU and the mains cable.
Do not use equipment that is damaged.
7.2 a
e
nnUal
lectRical
It is advisable to check the integrity of the insulation and protective earth on an annual basis, e.g. U.K. PAT
testing.
Do not use equipment that is damaged.
7.3 g
c
eneRal
leaning
The camera body can be cleaned with a soft cloth and dampened by water or glass cleaner.
Never spray liquids directly on the camera; apply cleaning solution to the cloth, then wipe the camera body with
the dampened cloth.
Do not use abrasive or other detergents to clean the camera.
7.4 c
leaning tHe
At some point, it may become necessary to clean debris that may have settled on the CCD imaging sensor
window. Cleaning the camera window can provide effective results providing you carefully follow these
step-by-step directions.
WARNINGS
Only open the shutter (if present) using camera control software.
Do not attempt to manually open the shutter (if applicable) with your fingers or other objects.
The glass is located very close to the CCD imaging sensor. Scratching the window may require the
window to be replaced.
Cleaning the window is done entirely at the customers risk, window replacement is not covered under
the warranty. It's better to under-clean than to over-clean.
Version 1.4 rev 13 Oct 2017
s
c
afety
Hecks
& d
econtaMination
c
W
aMeRa
indoW
i
nfoRMation
35