Power Supply Unit; Power Requirements - Oxford Instruments ANDOR Newton CCD Hardware Manual

Table of Contents

Advertisement

Newton
2.3 p
s
oWeR
Upply
The standard PS-24 PSU is fitted with an IEC connector for the electrical supply input.
The connection to the Newton is made via a 5-pin DIN socket:
Pin
1
2
3
4
5
Shell
The optional PS-25 PSU can be used to achieve maximum cooling and has 2x settings, STANDARD COOLING or DEEP
COOLING. The connection to the Newton is also made via a 5-pin DIN socket.
2.4 p
R
oWeR
eqUiReMents
Ensure that the power connector for the Newton camera is inserted securely. The connector is keyed to
aid correct orientation.
Power Input: 100 - 240V, 50 - 60Hz, 1.6A.
The electrical mains lead should be certified for use in your country and in applicable countries the plug
must be fitted with a 240V 5A fuse.
If users use any other power supply they do so at their own risk.
Version 1.4 rev 13 Oct 2017
U
nit
PS-24
+5V DC 5A max.
N/C
Common
+15V DC 2A max.
-15V DC 0.8A max.
GND
PS-25
+5V DC 3.5A
7.5V DC 4A
Common
+15V DC 100mA
-15V DC 100mA
GND