Preventive maintenance
Preventive maintenance-kit
Only original parts from Siemens must be used in the
Servo... System. Spare parts and maintenance kits can
be ordered from your local Siemens representative.
When performing this maintenance, a "Preventive
maintenance-kit for Servo..." should be used.
The following parts shall be replaced and they are in-
cluded in the "Preventive maintenance-kit for Servo...":
A. Filters for the gas modules
B. Bacteria filter for the inspiratory pressure
transducer
C. Bacteria filter for the O
D. Nozzle units for the gas modules
Performing the Preventive maintenance
• Prepare the unit as described in section
"Preparations" in this chapter.
Inspiratory section
• Loosen the screw (1) and lift off the inspiratory
section cover (2).
Preventive maintenance
cell
2
7 - 3