Preventive maintenance
7
7 - 2
General
A "Preventive maintenance" must be performed at
least once every year as long as the unit is not used
more than normal. Normal operation is estimated to
correspond to approx. 5.000 hours of operation.
In some parts of the Preventive maintenance, as
described in this chapter, the software Field Service
System (FSS) is a requirement.
Some optional equipment used in the Servo... System,
but not covered by this Service Manual, may also
demand maintenance actions. Refer to the equipments
documentation. These units can e. g. be:
• Humidifier equipment.
• Other optional equipment.
Preparations
• Make sure that the Servo... System is properly
cleaned before performing any maintenance; routine
cleaning as well as extended cleaning. For cleaning
procedures, refer to the "Servo... Ventilator System –
User's manual".
• Make sure that the Servo... System works properly
before performing any maintenance.
• Set the On/Off switch on the User Interface to Off.
• Disconnect the mains power cable.
• Disconnect the gas supplies (wall and/or cylinder).
• Remove patient tubing.
• Remove Servo Guard bacteria filter from the
expiratory inlet.
Equipment
• Standard service tools.
• Barometer (or information about the actual barometric
pressure).
• Field Service System (FSS ). Recommended but not
required.
• Preventive maintenance-kit containing all parts
needed during the maintenance.
Siemens-Elema AB
Servo...