Siemens Servo i Service Manual page 20

Table of Contents

Advertisement

Description of functions
Nozzle unit
The plastic N
U
OZZLE
The valve diaphragm, controlled by the I
S
, regulates the gas flow through the G
OLENOID
M
.
ODULE
The complete plastic nozzle unit must be replaced
during the "Preventive maintenance".
After replacement, allow the diaphragm to settle
during approx. 10 minutes before gas pressure is
connected to the G
AS
Inspiratory solenoid
The gas flow through the G
3
the I
S
NSPIRATORY
OLENOID
The current supplied to the solenoid is regulated so
that the gas module will deliver a gas flow according
to the settings on the User Interface.
Gas module key
The G
M
are provided with a mechanical key
AS
ODULES
to prevent that the module is mounted in the wrong
slot.
The key consists of a plastic guide mounted under-
neath the module and a corresponding guide
mounted in the patient unit.
Connector muff
The C
M
ONNECTOR
UFF
to the I
P
NSPIRATORY
IPE
Inspiratory pipe
The I
P
NSPIRATORY
IPE
M
to the I
UFF
NSPIRATORY
The I
P
comprises:
NSPIRATORY
IPE
• Housing and locking lever for the O
bacteria filter.
• Housing for the S
AFETY
• Connection for measurement of inspiratory
pressure.
The pipe is provided with internal flanges with the
purpose to improve mixing of O
3 - 4
contains a valve diaphragm.
NIT
NSPIRATORY
M
.
ODULE
M
is regulated by
AS
ODULE
via the N
U
.
OZZLE
NIT
connects the G
M
AS
ODULE
inlet.
leads the gas from the C
ONNECTOR
O
.
UTLET
C
with its
ELL
2
V
.
ALVE
and Air.
2
O
cell
2
The O
C
is mounted in a housing on the I
ELL
2
P
and is protected by a bacteria filter.
IPE
AS
Maintenance including exchange of bacteria filter
according to the User´s manual. The bacteria filter
must also be replaced during the "Preventive
maintenance".
The O
cell gives an output voltage proportional to the
2
partial pressure of oxygen inside the Inspiratory pipe.
At constant ambient pressure this output is
proportional to the O
each O
cell, the output signal will stay at a fairly
2
constant level usually within 10–17 mV in normal air
and at standard barometric pressure during the life
time of the cell. The cell should be replaced when the
output level has started to decrease. In this situation,
repeated calibration (i. e. Pre-use check) of the O
concentration is required. The life time of the cell is
affected by the O
concentration (at the cell) in % and expected cell life
time in hours the following applies at 25
O
Conc. x Expected cell life = 500 000% hours.
2
The O
cell is automatically calibrated each time a Pre-
2
use check is performed (if O
ventilator).
An ID PROM is integrated into each O
information and remaining lifetime can be read by the
Servo... System.
Temperature sensor
outlets
A T
EMPERATURE
on top of the O
temperature inside the I
The output signal, corresponding to the temperature
in the I
NSPIRATORY
I
F
.
NTERNAL
AN
Inspiratory pressure tube
The I
NSPIRATORY
P
with the I
IPE
NSPIRATORY
A bacteria filter protects the pressure transducer on
PC 1781 P
RESSURE
Maintenance including exchange of bacteria filter
according to User´s manual. The bacteria filter must
also be replaced during the "Preventive
maintenance".
Siemens-Elema AB
concentration in percent. In
2
concentration. With a
2
is connected to the
2
cell. Its ID
2
S
is integrated into the connector
ENSOR
C
. This sensor measures the
ELL
2
S
NSPIRATORY
ECTION
S
, is used for regulation of the
ECTION
P
T
connects the I
RESSURE
UBE
P
T
RESSURE
RANSDUCER
T
from contamination.
RANSDUCER
E382 E407E 061 01 01 02
Servo...