Siemens MULTI-ECHO EPI 4A21 Installation And User Manual
Siemens MULTI-ECHO EPI 4A21 Installation And User Manual

Siemens MULTI-ECHO EPI 4A21 Installation And User Manual

Siemens scanner user manual

Advertisement

Works-in-Progress package
Version 2.0
For the
SIEMENS Magnetom under VA21B
Installation and User's Guide
NUMARIS/4A21
April 4th, 2003
Section of Medical Physics, University Hospital Freiburg, Germany
Contact:
Oliver Speck PhD, Hugstetterstr. 55, D 79106 Freiburg,
email: oliver.speck@uniklinik-freiburg.de, phone: +49 761 2703832

Advertisement

Table of Contents
loading

Summary of Contents for Siemens MULTI-ECHO EPI 4A21

  • Page 1 Works-in-Progress package Version 2.0 For the SIEMENS Magnetom under VA21B Installation and User’s Guide NUMARIS/4A21 April 4th, 2003 Section of Medical Physics, University Hospital Freiburg, Germany Contact: Oliver Speck PhD, Hugstetterstr. 55, D 79106 Freiburg, email: oliver.speck@uniklinik-freiburg.de, phone: +49 761 2703832...
  • Page 2: Table Of Contents

    Multi-Echo EPI Basic principles ... 3 Principle of Multi-Echo EPI ... 3 Image contrast of Multi-Echo EPI ... 4 Processing options ... 4 References ... 4 Software Installation Procedure ... 5 Installation ... 5 Sequences and Protocols... 7 General Remarks ... 7 New Parameters...
  • Page 3: Basic Principles

    Multi-Echo EPI 1 Basic principles 1.1 Principle of Multi-Echo EPI The basic timing scheme of Multi-Echo EPI is shown in Figure 1. It is based on the original EPI sequence proposed by Mansfield (1). After a single RF-excitation, n complete EPI images are acquired in a single shot. The phase gradient is rewound to the original starting position after each echo-image to ensure identical k-space trajectories for all echo-images (2).
  • Page 4: Image Contrast Of Multi-Echo Epi

    Multi-Echo EPI 1.2 Image contrast of Multi-Echo EPI The signal intensity of the single echo images of the Multi-Echo EPI sequence is simply given by: where S0 is the signal intensity at echo time 0, which corresponds to the proton density for long repetition times TR, and TE is the echo time.
  • Page 5: Software Installation Procedure

    Multi-Echo EPI 3. O. Speck, T. Ernst, L. Chang. Bi-exponential modeling of multi gradient echo data of the brain. Magn. Reson. Med. 45(6), 1116-1121 (2001) 4. O. Speck, J. Hennig. Motion Correction of Parametric fMRI Data from Multi- Slice Single-Shot Multi-Echo Acquisitions, Magn. Reson. Med. 46(5): 1023- 1027 (2001) 2 Software Installation Procedure 2.1 Installation...
  • Page 6 Multi-Echo EPI Page 6 of 9 Numaris/4 VA21B...
  • Page 7: Sequences And Protocols

    Multi-Echo EPI 3 Sequences and Protocols 3.1 General Remarks Since the sequence is derived from the standard EPI sequence, most of the user interface and features are identical to this method. Due to the large number of images created in a short time, the images are saved to the database in the mosaic format. One mosaic-image contains all slices of one echo time for one repetition.
  • Page 8 Multi-Echo EPI Separate Correction: In the source code of the sequence card a new parameter ‘Separate Correction’ has been introduced. If this parameter is enabled (during compilation) each echo image is preceded by a new set of navigators used for the Nyquist ghost correction by the reconstruction.
  • Page 9: Remarks On Reconstruction Performance

    Multi-Echo EPI t-Test: The online t-test and display work as in the original EPI sequence. However, the t-test only works properly when motion correction is selected! Alternatively, T2*-Maps can be processed within the BOLD-card. 3.3 Remarks on reconstruction performance The number of images that can be acquired with the Multi-Echo EPI method is very high (up to 30 images per second for 64*64 matrix).

This manual is also suitable for:

Numaris/4a21Va21b

Table of Contents