4
Verification

Operator's Check for Local Operation

Recommended test equipment

Table 4-1
troubleshooting. Equipment other than the recommended models can be used, provided minimum
specifications are satisfied.
Table 4-1
Instrument type
Digital voltmeter
Attenuators (2 required)
Switches (2 required)
[a] P = Performance, T = Troubleshooting
Procedure
1 Configure the switching system as illustrated in
Chapter 5, "Local
Figure 4-1
2 Once configuration is completed, press and depress the 10 numbered pushbuttons on the
driver front panel. Each LED should alternate between off and on as each key is pressed.
3 In addition, if any switching devices is connected (attenuators, relays, or switches), an audible
click should be heard from the unit actuated. Pressing any numbered pushbutton should not
cause any other pushbutton to change state.
42
lists the test equipment required for performance test verification and equipment
Recommended test equipment
Critical specifications
0 to +30 Vdc
Programmable, 4 sections
+5 Vdc, +15 Vdc, or +24 Vdc drive source
Operations".
Switching system for verification
Figure 4-1
by following all steps described in
Keysight 11713B/C Operating and Service Manual
[a]
Use
T
P, T
P, T