A&D SW Series Instruction Manual page 44

Super washdown scales
Table of Contents

Advertisement

Example of connection
RS-422
Host computer
SW
Address 01
SW
Address 02
SW
Address 16
 The polarity (A, B) of the host computer signal depends on the computer model.
Check the technical manual of the computer.
Data format
The data format for the RS-422/485 is the same as the RS-232C except the following:
 When used with the function setting "
different address to each scale using the function setting "
 Start all commands with "@##" (## is the address of the scale to send a command).
All replies from the scale start with "@##".
The format after "@##" is the same as the RS-232C, both in commands and replies.
 When used with the RS-485 interface (function setting: "
• When sending commands continuously, leave an interval of 500 ms or more
between commands.
• Do not use stream mode (function setting: "
Commands will not be received correctly and will be invalid.
Terminator (may be
built in the computer).
RDA
RDB
SDA
SDB
SDA
SDB
RDA
RDB
TRM
SDA
SDB
RDA
RDB
TRM
SDA
SDB
RDA
RDB
TRM
Host computer
SW
Address 01
SW
Address 02
SW
Address 16
Connect TRM to RDB
at the farthest SW from
the host computer.
5if 1
" (RS-422) or "
5if 2
prt 0
") when sending commands.
44
RS-485
Terminator (may be
built in the computer).
A
B
SDA
SDB
RDA
RDB
TRM
SDA
SDB
RDA
RDB
TRM
SDA
SDB
RDA
RDB
TRM
5if 2
" (RS-485), set a
adr
##". ((##=01 to 99)
"), note the following: