Rcb 200Nx - Ametek 200 Series Quick Start Manual And Safety Instructions

Automotive transient immunity test generators
Table of Contents

Advertisement

AMETEK CTS
4.3.

RCB 200Nx

1
BAT input 13.5V ( Power supply RCB 200N1)
2
Fuse 20A
3
Ext. Trigger mode
4
Indication Relays EOL (end of lifetime)
5
Transient Generator switching
6
Pulse C Output CI 220, CI 260
7
Pulse A, F Output CI 220, CI 260
8
Pulse Output RI 130 (BNC)
1
Battery input
Power supply input for RCB200N1, nominal 13.5Vdc (max 18Vdc)
2
Fuse
Fuse 20A for device protection
3
Ext. Trigger mode
Indication of external Trigger mode. The controller detect when the bridge between Trigger OUT – Trigger In is
disconnected. In this Mode the user has to connect an external trigger signal for relays switching.
4
Relays EOL End of Lifetime indication
The lifetime of the Potter & Brumfield relays is limited by the erosion during BURST testing. The operating
duration is limited to 100hours.
- LED blinking:
- LED on: The test will be stopped. The user must replace the relays.
5
Transient generator switch settings
The selected pulse the Ford standard recommends to set different switches. The LED indicates the internal
relays setting according the Ford Standard
6
Pulse C output to DUT
Output plugs for testing pulses C-1 and C-2 as per CI 220
7
Pulse A, F output to DUT
Output plugs for testing pulses A1, A2-1, A2-2 and F as per CI 220 and CI 260
8
Pulse output to clamp
Output BNC plugs for testing as per RI 130. The BNC plugs are internal shorted when RI130 is not selected.
Lifetime is within 10 hour of max liftime
Automotive Transient Immunity Testing
9
START / STOP ; LED endless / sequence
10
MODE selection
11
PULSE Transient pulse selection
12
ID Test selection
13
TEST ON
14
Internal Trigger OUT
15
Trigger IN
1.00
Figure 2.1
17 / 26