Modbus holding registers
A.7 External input
A.7
External input
Table A- 14
External input, MAG 6000
MODBUS
MODBUS
register
address
4:02301
2300
4:02302
2301
4: 02303
2302
Table A- 15
External input, MASS 6000
MODBUS
MODBUS
register
address
4:02301
2300
4:02302
2301
4: 02303
2302
50
No. of bytes Data type
2
Byte
2
Byte
4
Float
No. of bytes Data type
2
Byte
2
Byte
4
Float
Description
Input Function:
0 = Off
•
1 = Start Batch
•
2 = Hold/Continue
•
3 = Stop Batch
•
4 = Zero Adjust (only SV)
•
5 = Totalizer reset
•
6 = Force output
•
7 = Freeze output
•
8 = Qmax 2 (night)
•
9 = Manual cleaning
•
Totalizer reset:
0 = Totalizer 1
•
1 = Totalizer 2
•
2 = Totalizer 1 + 2
•
Force output (1/100%)
Description
Input Function:
0 = Off
•
1 = Start Batch
•
2 = Hold/Continue
•
3 = Stop Batch
•
4 = Zero Adjust
•
5 = Totalizer reset
•
6 = Force output
•
7 = Freeze output
•
Totalizer reset:
0 = Totalizer 1
•
1 = Totalizer 2
•
2 = Totalizer 1 + 2
•
Force output (1/100%)
Operating Instructions, 06/2010, A5E00753974-04
Read / write
R/W
R/W
R/W
Read / write
R/W
R/W
R/W
Modbus RTU RS485