Functions Operation Table; Functions Indication Table - IDEAL 61-517 GFCI Operation And Safety Manual

Receptacle/voltagetester
Table of Contents

Advertisement

Available languages

Available languages

Functions Operation Table

Button

Functions Indication Table

Function
LCD
LCD Backlight White backlight.
High Voltage
Alarm
Over Range
Indication
Low Battery
Indication
APO
14
Response
Turns ON
Silent Mode
RED LED
Description
300 count LCD. Displays VAC and
1) Effective setting: ACV.
2) In ACV/DCV, voltage measurement value:
ACV/DCV ≥30V, LCD displays high voltage
alarm is
continuous tone.
LCD displays "OL" when over range is
encountered.
The "
with insufficient voltage. The low battery icon will
flash for 5 seconds and the unit will power off.
This ensures accurate readings.
No button presses for 3 minutes will be automat-
ically cause the unit to shut down and enter the
low-power state.
Default
Function
1) Press the on button for
OFF
longer than 2 sec. The unit
will turn on and be in the
silent mode.
2) Press the ON button for
Audible
less than 1 sec. The unit
will turn on and be in the
audible mode.
3) Press the ON/OFF
button for longer than
2 seconds when unit is
on to turn off. Unit will
OFF
automatically power off
after 2 minutes.
4) Press the GFCI Button
to test a GFCI.
Simultaneously the buzzer emits a
" symbol appears in the LCD display
Operation