Operating instructions Midwest Stylus Plus / Midwest Tradition
67 27 148  D3673.201.09.01.02
Temperature:
Overpressure:
Wrapping
Steam sterilization
Wrapped
Gravitation process
Wrapped
Gravitation process
Wrapped
Pre-vacuum process
Wrapped
Pre-vacuum process
NOTE!  Do not exceed 140 °C (284 °F), even during the drying
phase.
132 °C (270 °F) / 135 °C (275 °F)
1.87 bar (27.12 psi) / 2.13 bar (30.89 psi)
Holding time
15 minutes at 132 °C
10 minutes at 135 °C
4 minutes at 132 °C
3 minutes at 135 °C
7 Reprocessing
Drying time
30 minutes
30 minutes
30 minutes
16 minutes