SCT-3300 Signal Conditioning Transmitter Installation Manual March 11, 2020 PN 190937 Rev A...
Page 2
All information contained within this publication is, to the best of our knowledge, complete and accurate at the time of publication. Rice Lake Weighing Systems reserves the right to make changes to the technology, features, specifications and design of the equipment without notice.
5.0 Specifications ................8 Technical training seminars are available through Rice Lake Weighing Systems.
Page 4
SCT-3300 Rice Lake continually offers web-based video training on a growing selection of product-related topics at no cost. Visit www.ricelake.com/webinars www.RiceLake.com Visit our website...
Do not operate or work on this equipment unless this manual has been read and all instructions are understood. Failure to follow the instructions or heed the warnings could result in injury or death. Contact any Rice Lake Weighing Systems dealer for replacement manuals.
SCT-3300 Installation Load Cell Connections Connect up to four load cells. The load cells are summed and provide an analog output signal. Adjustments for zero and span can be made. There is an adjustment to increase stability to reduce the effect of environmental vibration.
5. If the value in mV is lower than the calculated value move on to the Load Cell Verification. 6. If the value in mV is greater than the calculated value move on to the Load Cell Verification. If problems exist after completing the installation check, please call Rice Lake Weighing Systems for support. Note...
SCT-3300 Analog Output Modification J1 Jumper Coarse Span Fine Zero Fine Span Coarse Zero J2 Jumper Figure 2-2. Analog Output Modification • Remove the J1 jumper to obtain an analog output of 0–10 VDC (output on terminals 22–23) • Remove the J1 jumper to obtain an analog output of 0–20 mA (output on terminals 21–23) •...
Use a cell simulator and a voltmeter to display in 0.1 of a millivolt. 1. Connect the simulator to the SCT-3300. 2. Measure the mV input on the terminals 4 (+) and 3 (-) and verify 0 mV; verify the mA or VDC output is 0-4 mA or 0 VDC (in accordance with the model used) otherwise use the DIP switches of coarse zero and the fine zero to set the output to those values (usually the DIP switch 1 is in position ON).
0-20mA, 4-20 mA (terminals 21-23) Table 5-1. Technical Specifications The SCT-3300 complies with Part 15 of the FCC Rules. Operation is subject to the following conditions: • This device may not cause harmful interference. • This device must accept any interference received, including interference that may cause undesired operation.
Page 14
Specifications subject to change without notice. Rice Lake Weighing Systems is an ISO 9001 registered company. 230 W. Coleman St. • Rice Lake, WI 54868 • USA U.S. 800-472-6703 • Canada/Mexico 800-321-6703 • International 715-234-9171 • Europe +31 (0)26 472 1319...