Teledyne TIENet 301 Installation And Operation Manual page 23

Ph/temperature device
Table of Contents

Advertisement

301 pH - pH Calibration - 4 buffer
Enter exact buffer value:
301 pH - pH Calibration - 4 buffer
Place the probe in the pH buffer and
then select NEXT to continue.
Rinse the probe and submerge in
the buffer solution at least one inch
beneath the surface.
Note
The actual value displayed for "pH counts" is insignifi-
cant to the user. It is a function of the voltage-to-fre-
quency conversion transmitted from the 301 device and
read by the Signature flow meter as a pH value.
Figure 2-10 301 Calibration sequence
Configure Options
1. Site Setup
2. Measurement Setup
3. Adjust
4. Equation/Trigger Setup
Adjust Options
Level
pH (301)
pH Calibration
301 pH
X.XX
Select NEXT to continue.
TIENet® 301 pH/Temperature Device
Section 2 Installation and Setup for Signature
5. Data Storage/Push Setup
6. Sampler Setup
7. Inputs/Outputs/Alarms Setup
8. Reset Totalizers
9. Reports/History Setup
Perform pH Calibration of sensor
301 pH
Select pH calibration buffer or Done
1. 4 buffer
2. 7 buffer
3. 10 buffer
4. Done
1, 2, or 3
If the pH buffer is not the
exact value, e nte r t he
value printed on the side
of the container.
T h e s c r e e n w il l c y c le
repeatedly. Wait for the
n u m e r ic al v a l u e d is -
p l a y e d n e x t t o " p H
counts" to stabilize to
within ±10 counts, and
then press NEXT.
pH Counts:
Select NEXT when stable
(Difference less than 10)
pH Calibration Complete for sensor
301 pH
Place pH probe back into the flow.
pH measurements will resume in 5
minutes.
Select NEXT to continue.
4
XXXXXXX
2-11