Fitting Instructions; Securing The Fixture To A Solid Surface; Fitting The Transducer - norbar 80026 Operator's Manual

Hydraulic wrench calibration fixture

Advertisement

FITTING INSTRUCTIONS

Securing the Fixture to a Solid Surface

The fixture should be securely fastened to a solid bench of sufficient strength to withstand the weight of the
fixture. Provision should be made to machine 3 holes into the bench to accommodate the supplied M8 x 50
long cap head screws. Use the fixture as a guide for correctly positioning the holes. The M8 screws should be
tightened against M8 nuts and washers (not supplied).

Fitting the Transducer

Once the fixture has been secured to a solid surface the static transducer needs to be fitted to the fixture.
1.
Remove the square and round reaction posts from the top reaction plate.
2.
Remove the lifting eye bolts. Then remove the bearing cover by undoing the 3 countersink screws.
3.
Remove the bolts securing the reaction plate to the base.
NOTE:
For 80026, 80022 and 80023 only, screw down the jacking screws to lift the reaction
top plate away from the dowels.
4.
Once the reaction plate is removed the transducer needs to be placed into the female square in the
base of the fixture. Select the correct transducer to match the fixture (see previous table).
5.
Make sure the electrical connector socket is in line with the side hole and place the 'o' ring if required
onto the transducer.
6.
Drop the inner bearing sleeve onto the transducer and replace the reaction plate back onto the base.
Tighten up the bolts securing the reaction plate to the base, see specifications on page 4 for N·m values.
7.
Fit the bearing cover back into the reaction plate.
8.
Plug the electrical lead into the socket of the transducer through the side hole.