Specifications And Environmental Conditions; Specifications - Fluke Calibration CalPower 6331 User Manual

Calibration bath
Table of Contents

Advertisement

3
Specifications and Environmental
Conditions
3.1

Specifications

Range
Stability (2 sigma)
Uniformity
Heating Time
Cooling Time
Stabilization Time
Temperature Setting
Set-point Resolution
Display Resolution
Digital Setting Accuracy
Digital Setting Repeatability
Control Heater
Boost Heater
Access Opening
Depth
Wetted Parts
Power
System Fuse
Volume
Weight
Dimensions
Safety
Interface Package
Rated at listed 115 V (or optional 230 V)
35°C to 300°C
±0.007°C at 100°C (oil 5012)
±0.010°C at 200°C (oil 5017)
±0.015°C at 300°C (oil 5017)
±0.007°C at 100°C (oil 5012)
±0.017°C at 200°C (oil 5017)
±0.025°C at 300°C (oil 5017)
140 minutes, from 35°C to 300°C (oil 5017)
(with boost)
900 minutes, from 300°C to 100°C (oil 5017)
15–20 minutes
Digital display with push-button data entry
0.01°C; high-resolution mode, 0.00018°C
0.01°C
±1°C
±0.01°C
700 Watts
6.8" x 4.7" (172 x 119 mm)
18" (457 mm) without liquid-in-glass calibration kit
19" (482 mm) with liquid-in-glass calibration kit
304 Stainless Steel
115 VAC (±10%), 50/60 Hz, 14.8 A, 1750 W
[230 VAC (±10%), 50/60 Hz, 7.4 A, 1860 W
115V: 20A 250V fast blow
230V: 10A 250V fast blow
4.2 gallons (15.9 liters)
90 lb. (41 kg)
14" W x 31" D x 42" H (356 x 788 x 1067 mm), 37" (940 mm) from
floor to tank access opening
OVERVOLTAGE (Installation) CATEGORY II, pollution Degree 2 per
IEC 1010-1
RS-232 included, IEEE-488 optional

3 Specifications and Environmental Conditions

900 Watts
Specifications