Installing the AquaPrep
6. Fuse the power line according to local safety practices.
The double-fuse arrangement requires 20-mm fuses.
Fuse Block
The power cord must be disconnected from the degasser when installing or replacing
fuses. Failure to do so could result in electrical shock.
a. Observe the position of the fuse block, using the illustration shown above for
reference.
• If the single-fuse arrangement is desired, position the fuse block so that the side with
the single fuse slot and the jumper bar is away from the cover.
• If the double-fuse arrangement is desired, position the fuse block so that the side
with the double fuse slots is away from the cover.
b. If the fuse block is positioned properly for the desired fusing, proceed to Step c.
If the fuse block is not positioned properly:
1) Remove the fuse block retaining screw.
2) Lift the fuse block from the cover.
3) Rotate the fuse block.
4) Mount the fuse block to the cover.
5) Replace the retaining screw.