C MultiVolume Insert Option
1, 10, 3.5, 35, 650,
1. Remove the sample chamber cap and place on a clean work surface with the greased side
facing upward.
2. Place the insert and sample cup into the sample chamber. Replace the chamber cap.
3. Calibrate the insert. See
4. See
Verify Operation on page 10 -
O
P
PERATING
ARAMETERS
To operate using inserts, access the Chamber Insert prompt to select the appropriate MultiVolume
insert.
1. Press Alt + 2 to access the Setup function.
2. Press CHOICE until Report Options is displayed, then press ENTER.
3. Continue pressing ENTER until the Request Insert prompt is displayed.
4. Press CHOICE until Yes is displayed, then press ENTER.
5. Press SAVE to save the changes and return to the Reload prompt.
A Chamber Insert prompt will display during all normal operating procedures. For example, a
Chamber Insert prompt will display when starting a calibration or analysis.
C - 6
CM 3
1300
I
AND
NSERTS
About Calibration on page 10 -
12.
U
K
SING THE
1.
EYPAD
AccuPyc II 1345 Operator Manual