Connect a Gas Cylinder
Regulator Pressure Settings
Analyzer Series
3Flex
AccuPyc
ASAP
AutoChem
AutoPore
Gemini
MIC SAS II
TriStar
Exceeding the maximum recommended air pressure could cause personal injury or dam-
age the instrument.
AccuPyc II 1345 Operator Manual
134-42851-01 (Rev -) — Not Released
Gauge should indicate
15 psig (103 kPag)
25 psig (172 kPag)
15 psig (103 kPag)
15-18 psig (103 - 124 kPag)
50-60psig (345 - 404 kPag)
15-18 psig (103 - 124 kPag)
15-20 psig (103-138 kPag)
15-18 psig (103 - 124 kPag)
Replace a Gas Cylinder