R
P
C
ESET
RESSURE
Contact a Micromeritics Service Representative and schedule a proper pressure cal-
ibration since data accuracy may be compromised.
This option can be used when the pressure is suspected of not being reported correctly by the
analyzer. For example: a zero, negative, or unreasonably high reading is being consistently
reported. Resetting to nominal may restore normal pressure readings but a proper calibration should
be performed.
Display Text
Calibration Type?
Reset Pressure Cal
Reset Pressure cal?
[Yes / No]
AccuPyc II 1345 Operator Manual
134-42851-01 (Rev -) — Not Released
ALIBRATION
Description
Press CHOICE until the desired option is displayed, then press
ENTER.
Resets the pressure calibration to nominal values. Temperature and
volume calibrations are not affected. Nominal values may be sufficient.
Schedule a proper pressure calibration as soon as possible, then
press CHOICE.
Select Yes to reset the calibration to nominal values. Press ENTER.
Select No to contact your Service Representative to schedule a proper
pressure calibration. Press ENTER.
Reset Pressure Calibration
10 - 17