one standard is shipped with the 1-, 100-, and 2000 cm
l
two are shipped with the 10 cm
l
three are shipped with the 350 cm
l
10. Remove the sample chamber cap, place the sample cup back into the chamber, and replace
the cap. Turn the cap clockwise to tighten.
11. Allow the standards to reach thermal equilibrium with the pycnometer.
Wait 10 minutes for 1- and 10 cm
l
Wait 15 minutes for the 100 cm
l
Wait 20 minutes for the 350 cm
l
30 minutes for the 2000 cm
l
12. Click
Continue
to complete the second phase of the calibration (volume scale). Status mes-
sages display until the calibration is complete.
13. Click
Done
to accept the new values, then click OK.
14. Do not remove the standard from the sample chamber until the operation is verified. See
Verify Operation on page 10 -
AccuPyc II 1345 Operator Manual
134-42851-01 (Rev -) — Not Released
3
pycnometer
3
pycnometer
3
pycnometer
3
pycnometer
3
pycnometer
3
pycnometer
12.
Calibrate Volume Scale
3
pycnometers
10 - 9