Change The Ldo Or Lbod Settings - Hach HQ1110 User Manual

Hide thumbs Also See for HQ1110:
Table of Contents

Advertisement

Option
Temperature
Correction Factor
Acceptance range Sets the range of values that will be accepted for verifications. Options:
Invalidate
calibration
Reminder
Snooze reminders When Reminder is set to a time interval, lets the user ignore a verification
Snooze time
6. Select Probe settings to manage the probe settings as a group.
Option
Current
Create new
settings
Delete settings

9.4 Change the LDO or LBOD settings

Refer to the steps that follow to change the settings for measurements, calibrations and verifications
with an LDO or LBOD probe. The user can save the settings with a user-specified name.
1. Push
and select Settings. The list of settings shows.
2. Select LDO[probe number] or LBOD[probe number].
3. Select Measurement to change the settings for sample measurements.
Option
DO units
Resolution and
speed
Lower limit
Upper limit
Description
WhenStandard is set to Custom, sets the temperature correction factor for
a custom standard solution. Options: 0%/°C to 9.99%/°C (default:
1.90%/°C).
1 to 9% from the expected value (default: 8%).
Changes the calibration status to expired if the verification result is not
within the Acceptance range setting. Options: Yes or No (default).
Sets a time interval for verification. A reminder shows on the meter after
the time interval. Options: Off (default), 30 minutes, 2 hours, 4 hours,
8 hours, 12 hours or 24 hours.
reminder for a selected time. Options: Yes (default) or No.
When Snooze reminders is set to Yes, sets the time when the verification
reminder will show again. Options: 5 minutes, 10 minutes, 15 minutes
(default), 30 minutes or 60 minutes.
Description
Selects the probe settings to use for measurements, calibrations and
verifications. Options: Default, Edited or User-specified name. The Edited
option shows when the current settings have changes but are not saved with
a user-specified name.
Saves the current probe settings with a user-specified name (maximum
16 characters, letters and numbers).
Removes a user-specified probe setting that is not in use.
Description
Sets the measurement units. Options: mg/L (default) or %.
Sets the number of decimal points and stabilization time. A slow speed gives
more time for the measurement to become stable. Options: 0.1 - Fast -
(0.35mg/L)/min, 0.01 - Fast - (0.35mg/L)/min, 0.01 - Normal - (0.15mg/L)/min
(default) or 0.01 - Slow - (0.05mg/L)/min.
Sets the minimum accepted value for the measurement. Options: 0.00 to
19.90 mg/L or 0 to 199% (default: 0.00 mg/L).
Sets the maximum accepted value for the measurement. Options: 0.10 to
20.00 mg/L or 1 to 200% (default: 20.00 mg/L).
English 25