Guidance and Manufacturer's Declaration - Electromagnetic Immunity:
The AutoVector
Fluid Immersion Simulator is intended for use in the electromagnetic environment specified below.
®
The customer or the user of the Auto Vector Fluid Immersion Simulator should assure that it is used in such an
environment.
Immunity Test
IEC60601 Test Level
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/output
lines
Surge
±1 kV line(s) to
line(s)
IEC 61000-4-5
±2 kV line(s) to earth
Voltage
<5% Uτ
dips, short
(>95% dip in Uτ)
interruptions
for 0,5 cycle
and voltage
variations on
40% Uτ
power supply
(60% dip in Uτ)
input lines IEC
for 5 cycles
61000-4-11
70% Uτ
(30% dip in Uτ)
for 25 cycles
<5% Uτ
(>95% dip in Uτ)
for 5 sec
Power frequency
3 A / m
(50/60 Hz)
magnetic field
IEC 61000-4-8
Note: Uτ is the a.c. mains voltage prior to application of the test level.
Dolphin Fluid Immersion Simulation® Systems
Compliance
Electromagnetic Environment-Guidance
±6 kV contact
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
±8 kV air
relative humidity should be at least 30%.
±2 kV for power
Not applicable Mains power quality should
supply lines
be that of a typical commercial or hospital
environment.
Not applicable
±1 kV line(s) to
Mains power quality should be that of a typical
line(s)
commercial or hospital environment.
±2 kV line(s) to earth
<5% Uτ
Mains power quality should be that of a typical
(>95% dip in Uτ)
commercial or hospital environment. If the user
for 0,5 cycle
of the AutoVector Fluid Immersion Simulator
requires continued operation during power mains
40% Uτ
interruptions, it is recommended that the Auto
(60% dip in Uτ)
Vector Fluid Immersion Simulator be powered
for 5 cycles
from an uninterruptible power supply or a battery.
70% Uτ
(30% dip in Uτ)
for 25 cycles
<5% Uτ
(>95% dip in Uτ)
for 5 sec
Not applicable
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.