Appendix A: Electromagnetic Compatibility (EMC) Information
AutoVector
Fluid Immersion Simulator
®
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided in this manual.
Portable and mobile RF communications equipment can affect Medical Electrical Equipment.
The use of Accessories, transducers, and cables other than those specified by the manufacturer, may result in
increased Emissions or decreased Immunity of the Auto Vector Fluid Immersion Simulator.
The AutoVector Fluid Immersion Simulator should not be used adjacent to or stacked with other equipment and that if
adjacent or stacked use is necessary, the Auto Vector Fluid Immersion Simulator should be observed to verify normal
operation in the configuration in which it will be used.
Guidance and Manufacturer's Declaration –Electromagnetic Emissions
The AutoVector Fluid Immersion Simulator is intended for use in the electromagnetic environment specified below.
The customer or the user of the Auto Vector Fluid Immersion Simulator should assure that it is used in such an
environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
16
© 2014 Joerns Healthcare • 6110109 RevG • 14-2640
Compliance
Electromagnetic environment – guidance
The AutoVector Fluid Immersion Simulator uses RF energy only for its
Group 1
internal function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
Class B
The AutoVector Fluid Immersion Simulator is suitable for use in all
Class A
establishments, including domestic establishments and those directly
connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Complies