Lamp Replacement

NOTE: If lamp surface is touched or becomes smudged, wipe clean with a soft
cloth. This will assure longer lamp life and higher output. Lamps can be very
hot. Be sure lamp is cool before touching. For guaranteed performance, use
only genuine Welch Allyn lamps.
NOTE: When adequately tightened, instruments will be able to be swiveled in
both directions. After changing a lamp, to achieve the ability to swivel in both
directions, press firmly with thumb on the knurled portion of the instrument
adaptor while giving the instrument a sharp twist in the clockwise direction.
Ophthalmoscope Lamp Replacement
1. Place thumb on knurled ring at the base
of the instrument. Unscrew ophthalmo-
scope head in a counterclockwise
direction.
2. Grasp end of lamp and pull out (use
nail file or similar object, if necessary).
3. Insert replacement lamp (#03900) by
carefully aligning pin on lamp shaft with
slot in ophthalmoscope head. Push lamp
in until firmly seated.
12/7/01
10:58 AM
Page 9
Otoscope Lamp Replacement
1. Place thumb on knurled ring at the
2. Grasp base of lamp and pull out of
3. Insert replacement lamp (#03400)
9
Align Pins
base of the instrument. Unscrew
scope head in a counterclockwise
direction and remove.
instrument.
by reversing above process.