Dräger Medical Narkomed GS Setup And Installation Manual page 28

Anesthesia system
Hide thumbs Also See for Narkomed GS:
Table of Contents

Advertisement

3
Setup and Installation
Installing the
1. Hold the locking bar on the anesthesia machine's fresh gas outlet in
Fresh Gas
the extended position.
Oxygen
Sensor
2. With the oxygen sensor port facing up, slide the notched portion of the
Adapter
adapter stem into the retaining slot of the locking bar, so that the
(Optional)
adapter's 15 mm male fitting faces the anesthesia machine's fresh gas
outlet.
3. Release the locking bar, guiding the adapter's 15 mm male fitting into
the fresh gas outlet of the anesthesia machine.
4. Insert the oxygen sensor into the port on top of the adapter.
WARNING:If the oxygen sensor is removed, the oxygen sensor port must
5. Attach the non-rebreathing circuit fresh gas hose to the adapter's
barbed fitting. (For non-rebreathing circuits equipped with a 15 mm
fresh gas connector, cut off the connector and press the hose onto the
barbed fitting of the adapter.)
FRESH GAS
OUTLET
OXYGEN
SENSOR
15MM MALE
FITTING
ADAPTER
STEM
OXYGEN
SENSOR PLUG
3-18
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
be closed with the oxygen sensor port plug.
BARBED FITTING FOR
NON-REBREATHING
Narkomed GS Setup and Installation Manual
RETURN TO CD-ROM TABLE OF CONTENTS
OXYGEN SENSOR PORT
FRESH GAS HOSE
GAS CIRCUIT
SU20997
FRESH GAS
LOCKING BAR
(EXTENDED POSITION)