Error Reporting - GE LOGIQ 180 Basic Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2317229, R
IRECTION
EVISION
Section 7-3

Error Reporting

7-3-1
Error Reporting
The following are the error messages available:
Table 7-22
Errors
"THE EVEN CHECKSUM FAILED"
"THE ODD CHECKSUM FAILED"
"THE RAM TEST FAILED"
"THE GRAPHICS MEMORY TEST FAILED"
"NVRAM TEST FAILED"
"TESTING FLASH BANK 0 ---- FAILED"
"NO FLASH ASSY CARD DETECTED"
"TESTING CINE DEVICE U3E1S --
"THE IMAGE MEMORY TEST FAILED "
"THE KEYBOARD SUBSYSTEM FAILED TO
RESPOND"
"KEYBOARD NOT CONNECTED"
"NO DSC FRAME FREEZE ACKNOWLEDGE" This message is displayed during system boot up when the DSC is failed.
"NO SCBF FRAME FREEZE
ACKNOWLEDGE"
"SCAMP MOUSE CONTROLLER SELF TEST
ERROR"
"TRACKBALL INTERFACE ERROR"
"NO TRACKBALL ACKNOWLEDGEMENT"
"NO SCBF SYSTEM MODE ACKNOWLEDGE" This message is displayed when the system boots up in the normal mode. when FEB is failed.
"NO SCBF SCAN MODE ACKNOWLEDGE"
"CINE HARDWARE NOT PRESENT"
CINE HARDWARE DIAGNOSTICS FAILED"
"NO CINE HARDWARE TO SUPPORT 64
FRAMES"
CINE HARDWARE SUPPORTS 32 FRAMES
ONLY"
"PLEASE POWER OFF AND RESTART THE
SYSTEM! "
"PRINTER IS BUSY"
"PRINTER ERROR"
"PRINTING ABORTED"
3
Error Messages
This Message comes up during the diagnostics of the CPU Board. When the Eprom checksum
(even) test for the CPU board fails, this message is displayed.
This Message comes up during the diagnostics of the CPU Board. When the Eprom checksum
(odd) test for the CPU board fails, this message is displayed.
This Message comes up during the diagnostics of the CPU Board. When the R/W memory test
for the CPU board fails, this message is displayed.
This Message comes up during the diagnostics of the CPU Board. When the Graphics memory
test for the CPU board fails, this message is displayed.
This Message comes up during the diagnostics of the CPU Board. When the NVRAM test for
the CPU board fails, this message is displayed.
This Message comes up during the diagnostics of the CPU Board. When the Flash memory Bank
test for the Flash memory fails, this message is displayed.
This Message comes up during the diagnostics of the CPU Board. When the Detect Flash Assy
test for the Flash memory fails, this message is displayed.
FAILED This Message is displayed during the diagnostics of the DSC board. When the cine memory test
for the DSC board fails, this message is displayed. U3E1S may be different depending on the
name of the cine device.
This Message is displayed during the diagnostics of the DSC board. When the image memory
test for the DSC board fails, this message is displayed.
This Message is displayed during the diagnostics of the Keyboard. When the keyboard firmware
is not alive, this message is displayed.
If the keyboard is not connected then this message is displayed during system boot up.
This message is displayed during system boot up when FEB is failed.
This message is displayed during system boot up when CPU is failed.
This message is displayed during system boot up and the trackball interface to keyboard is not
proper.
This message is displayed during system boot up and keyboard connection is not there.
This message is displayed during system boot up when FEB is failed.
This message is displayed during system boot up and cine hardware is not present or some error
is there.
This message comes up during the system boot up, when the DIP Switch is set for 64 frames
and cine hardware for only 32 frames are connected.
This message is displayed when the system boots up in the normal mode and any error which
needs the system to be restarted comes up.
This message comes up when shift + record is pressed without connecting the printer.
This message comes up when an error occurs in the printing process.
This message is displayed during Printing & due to printer error.