25. Basis Performance Test

To keep the quality of the unit, regular performance testing is
recommended. This section describes a method and specification of
basic performance testing. If a problem is found in the results of
basic performance testing, or formal testing is needed, please contact
the dealership where you purchased the product, or contact us.

25.1 Frequency characteristics

Adjust the output level of the spectrum analyzer calibration
unit (thereafter, „calibration unit") so that the displayed power
value is -15dBm at each frequency for this unit, and measure
the absolute value with a receiver for calibration (microwave
power meter, etc.).
Setting of HM5033
Centerfrequency
Frequency span RBW
50 kHz
100 kHz
1 MHz
10 MHz
100 MHz
1 GHz
2 GHz
3.3 GHz
RBW switching error is included at RBW other than 3MHz.
Subject to change without notice
200 kHz
10 kHz
within Reference ± 2.6 dB ± 1dot
200 kHz
30 kHz
within Reference ± 2.6 dB ± 1dot
2 MHz
100 kHz
within Reference ± 1.6d B ± 1dot
10 MHz
3 MHz
within Reference ± 1.0 dB ± 1dot
10 MHz
3 MHz
10 MHz
3 MHz
within Reference ± 1.0 dB ± 1dot
10 MHz
3 MHz
within Reference ± 1.0 dB ± 1dot
10 MHz
3 MHz
within Reference ± 1.0 dB ± 1dot
Specifications
Reference
Basic Performance Test
Measurement value
Judgement