Copyright Table of Contents Shenzhen Lepu Intelligent Medical Equipment Co., Ltd 1. Product Overview .........01 1.1 Appearance ..........01 Statement 1.2 Name and Model ........01 The company owns all the copyrights of this manual, 2. Intended Use ..........01 including published and unpublished documents, and 3.
The product includes both visual and audible alerts for absorbed red and infrared light with each pulse. high/low SpO2 and Pulse Rate. The applied part of the HbO-SMART is constructed from silica gel. Please carefully read the User Manual before use.
11. The device complies with electromagnetic compatibility requirements 4. Excessive patient movement may be erroneously identified as pulse for electronic medical products or systems in IEC60601-1-2. Radio signals and may affect the measurements of this device. transmission equipment or other electromagnetic interference may affect 5.
Not for continuous monitoring 6. Read measurement from the device screen. (no alarm for Spo2) 7. The HbO-SMART has 4 different user screen options. Once the screen. measurement is stable, pressing the power button will change the screen Pulse intensity bargraph...
8. Setting Menu Setting range Default setting Limit Lo 85%~99% PR Limit Lo 30bpm~100bpm 50bpm Figure 6 Figure 5 PR Limit Hi 100bpm~200bpm 120bpm Sound Volume Level 1~5 and Off can be selected Exit Long press to exit %SpO %SpO During a non-measurement condition“Finger out”, by pressing and holding the power button the user can enter the menu settings as shown below.
9.Lanyard Installation 11.Maintenance, Storage and Transportation 1. Pass the thinner end of the lanyard through the slot on the device as The life cycle of this device is 5 years when used daily for a number of 10 shown in Fig 12. measurements, for periods of 10 minutes each measurement.
Page 9
·Clinical testing is a commonly used method for determining oxygen 12.Technical Specifications saturation accuracy. The arterial hemoglobin oxygen saturation measured with the device should be compared with the result of sampled arterial blood analyzed with CO-oximeter. Display mode OLED ·The name of the simulator used is Index2 FLUKE simulator, version Measuring range 70%~99% number: 3.0.0.
(1) Anti-shock protection category Internal power supply Anti-shock protection measure Type BF applied part Guide and manufacturer's statement - Electromagnetic emissions Waterproof protection measure IP22 This product is intended for use in the following electromagnetic environment.The purchaser or user of theproduct should ensure that it Net Weight approx.
Page 11
(3) Guide and manufacturer's statement - Electromagnetic immunity ± 2 kV for power Electrical fast cord This device is intended for use in the following electromagnetic Not applicable Not applicable transient burst ± 1 kV for input/ environment. The purchaser or user of this device should ensure that it output lines is used in this electromagnetic environment: ±...
Page 12
(4) Recommended isolation distances between portable and mobile RF Where: communication equipment and this product P - The transmitter's maximum rated output power, in watts (W), provided by the transmitter's manufacturer; This product is intended for use in electromagnetic environments where d - Recommended isolation distance in meters (m) b.