Name
Type of signal Board
1
\EXT-SW
analog
2
\INTBON
digital
3
\V1-TRG
digital
4
\V2-TRG
digital
5
5V CPU
5V stabil.
6
ADD
digital
7
ALT-TB
digital
8
ANA_MODE
digital
9
ANA-Z
digital
10
AUTO_FOC
analog
11
BLANK
analog
12
BLOUT
digital
13
C_SW1
digital
14
C_SW2
digital
15
CB_SW
digital
I6
CS_AD
digital
17
CS_DA
digital
18
CS_MAIN
digital
19
CS1
digital
20
D_ANA
digital
21
DARK2
digital
22
DEL-EN
digital
23
DEL-POS
analog 10 bits MB
24
DIG_TR
digital
25
DY11
Signal
26
DY12
Signal
27
DY21
Signal
28
DY22
Signal
29
END-HO
digital
30
END-SW-A
31
END-SW-B
32
EXT-SW
Signal
33
FOC_CTR
analog
34
FOC_FC
analog
35
GATE-A
digital
36
GATE-B
digital
37
HOS
digital
38
IA*10
digital
39
IA*100
digital
40
IA*2
digital
41
IA*5
digital
42
I8*10
digital
43
IB*100
digital
44
IB*2
digital
45
I8*5
digital
46
INT A+B
analog 8 bits
47
INT A-B
analog 8 bits
48
INT RDOT
analog 8 bits
49
LF-TRG
digital
50
LINE-ON
digital
51
LINE-TRG
signal
52
MAG_10
digital
53
MAIN_CLK
digital
54
PP-
analog
55
PP+
analog
56
PP-DIS
digital
57
SCLK
digital
58
SDI
digital
59
SDO
digital
60
SH-DI
digital
Subject to change without notice
Meaning
MB, AT
signal from external trigger unit
MB
control signal for timebase B intensity
MB
trigger mode switch CH1
MB
trigger mode switch CH2
MB. FC
supply voltage for ADC, DACs, CPU
MB
addition CH1 + CH2
MB
alternate timebase
MB
controls analog- / digital mode
MB, RO
control signal for READOUT intensity (analog-/ digital intens.)
MB, CRT
optimizes beam sharpness (max./min. intensity)
MB, CRT
blanking signal
MB
unblank pulse
MB
selects timebase A timing and holdoff capacitors
MB
selects timebase A timing and holdoff capacitors
MB
selects timebase B timing capacitors
MB
chip select ADC
MB
chip select DACs
MB
chip select X-PLD, TB & TRG-register
MB
chip select V-PLD
MB, RO, YFA
controls digital-/analog readout
MB. RO
unblank signal for readout
MB
enable DELAV-comparator
DELAV-comparator reference voltage
MB
prepared for digital board
MB
prepared for digital board
MB
prepared for digital board
MB
prepared for digital board
MB
prepared for digital board
MB
indicates the end of the holdoff time
MB
indicates the end of timebase A ramp
MB
indicates the end of timebase B ramp
MB, AT
external trigger signal
MB, CRT
line from focus potentiometer to CRT
MB, FC
line from focus potentiometer to CRT
MB
timebase A control line
MB
timebase B control line
MB
minimum holdoff in AUTOSET condition
MB
timebase A
MB
timebase A
MB
timebase A
MB
timebase A
MB
timebase B
MB
timebase B
MB
timebase B
MB
timebase B
MB
intensity for A+B
MB
intensity for A-B
MB
intensity for READOUT
MB
control signal for LF trigger coupling
MB
control signal for line (mains) triggering
MB, PS
line (mains) trigger signal
MB
control signal for magnification
MB
gated SCLK for X-PLD & register
MB
negative peak value
MB
positive peak value
MB
control signal for peak value discharge (i.e. Autoset..)
MB, FC, RO
system clock
serial data ADC,RO → CPU
MB, FC, RO
serial data CPU → V-PLD. ADC. DAC. Register, RO
MB. FC. RO
serial data register → X-PLD
MB