• Device Address: If your Smart Interface is part of a Network, enter the Network Address here. The default network
Note:
The default Device Address is 1.
• Device IP or Port: The COM port number that your device is connected to on your computer.
Important:
The following parameters should NOT be changed. These settings should NOT be changed unless the
configuration has been done on the interface.
• BaudRate: Controls bits per second
• DataBits: The number of 'bits' in each character sent.
• Parity: A means of checking correctness of character by adding an extra 'bit' to the character and setting the value
based on all the other bits in the character.
• StopBits: The number of 'bits' used to indicate the end of the character.
Once you have completed setting the communication parameters for your device, click Finish.
4.3

Load Cell Interface

The SP-010 measures the bridge voltage to determine the applied force using bridge characteristics provided by the user. To
configure the settings on your SP-010, follow these instructions:
Step 1:
Click the Input Configuration Tab in SYNC and select the Load Cell input type form the Type drop down.
Step 2:
Set the Parameters of your Load Cell to your preferred settings.
If the Bridge Impedance is set to 0 the SP-010 will automatically calculate the impedance on the next device reset or power
cycle.
address is 1 for most devices. Please refer to the manual of your Smart Interface for more
information.
Parameter
mV/V
Full Scale
Impedance
TARE (Target)
Description
Bridge output at full scale
Bridge maximum load
Bridge Impedance
Update this parameter to TARE bridge
9 |
P a g e