Download Print this page

Miele PG 8583 Operating Instructions Manual page 96

Laboratory glassware washer

Advertisement

Program settings
Program header
96
Adjusting program settings
The program settings should be adjusted to suit technical
requirements and the load items.
Additional specialist knowledge is required to alter program settings
and this should therefore be undertaken only by experienced users or
by Miele Service.
Changing program parameters on a validated machine will
necessitate a renewed performance validation.
Program structure
Each program is subdivided into program blocks which run one after
another. A program consists of at least one and a maximum of 11
program blocks. Each block can occur only once in a program.
The so-called program header is placed above the program blocks
and contains general program settings. Individual wash block
parameters are also globally activated or deactivated here.
Spray arm monitoring
-
It is possible to monitor spray arm rotation in selected wash blocks.
Change water volume
-
The water intake quantity can be increased or reduced in each
program. The setting is then valid for all program blocks including
water intake.
Drainage time
-
If the on-site drainage system is insufficient to drain the waste
water from the wash cabinet within the time allocated, the drainage
time can be increased.

Advertisement

loading

This manual is also suitable for:

62858371cdn